Show Hamamatsu Avalanche Photo Diode 1501016501 - Archived data from Thu, 23. June 2016 10:41:01 CEST
This is all the information about APD 1501016501. If it is wrong, edit the data.
Subdetector specification: | |
---|---|
Serial: | 1501016501 |
Type: | Hamamatsu Avalanche Photo Diode |
Detector: | unassigned |
Unit: | unassigned |
Preamp: | 0 |
Current location: | unknown |
Installation information: | |
Label: | none |
Manufacturer information: | |
Wafer position: | G05 |
Break-through voltage: | 419 V |
Voltage for Gain 100 (T=+25°C): | 390.7 V |
Dark current: | 3.5 nA |
Screening Logistics: | |
Available: | Yes |
Storage Box: | 34 |
Position in Box: | 6 |
EP1 batch: | none |
EP1 batch after irradiation: | none |
Shipment: | |
Grid number: | none |
Position in grid: | none |
Arrival for irradiation: | none |
Sent for analysis after irradiation: | none |
Return for assembly: | none |
Irradiation: | |
Date: | none |
Dose used: | none |
Temperature: | none |
Position: | none |
Bias voltage: | none |
Annealing: | |
Date: | none |
Temperature: | none |
Duration: | none |
Measurement results: | |
Voltage for Gain 100: | T = +20 °C: none T = -25 °C: none |
Voltage for Gain 150: | T = +20 °C: none T = -25 °C: none |
Voltage for Gain 200: | T = +20 °C: none T = -25 °C: none |
Gain/Voltage slope at M = 100: | T = +20 °C: none T = -25 °C: none |
Gain/Voltage slope at M = 150: | T = +20 °C: none T = -25 °C: none |
Gain/Voltage slope at M = 200: | T = +20 °C: none T = -25 °C: none |
Break-through voltage: | T = +20 °C: none T = -25 °C: none |
Notes: |
Version history
Time | Author | Change comment |
---|---|---|
15. Nov 2022 12:51:38 CET | oafedulidis | Assigned to capsule by script |
01. Dec 2021 12:56:57 CET | oafedulidis | Updated box and position via API. |
05. Feb 2021 10:52:12 CET | jreher | Updated box and position via API. |
03. Feb 2021 18:18:19 CET | jreher | Updated box and position via API. |
11. Aug 2020 17:25:09 CEST | jreher | Updated APD characteristics via API. |
28. Nov 2019 12:22:36 CET | jreher | Updated APD location via API. |
25. Nov 2019 17:13:32 CET | jreher | Updated box and position via API. |
13. Jun 2019 11:29:14 CEST | jreher | Updated APD location via API. |
12. Jun 2019 16:47:55 CEST | jreher | Updated APD batch assignment via API. |
12. Jun 2019 16:32:50 CEST | jreher | Updated APD batch assignment via API. |
26. Nov 2018 14:59:52 CET | markus.moritz | |
29. Sep 2017 13:46:42 CEST | markus.moritz | Entered arrival for irradiation date. |
01. Aug 2016 17:29:42 CEST | Tobias | Assigned to grid via information extracted from APD DB. |
23. Jun 2016 10:41:01 CEST | Tobias | Imported from Hamamatsu datasheet. |