Show Hamamatsu Avalanche Photo Diode 1414016136 - Archived data from Tue, 11. August 2020 17:58:44 CEST
This is all the information about APD 1414016136. If it is wrong, edit the data.
Subdetector specification: | |
---|---|
Serial: | 1414016136 |
Type: | Hamamatsu Avalanche Photo Diode |
Detector: | unassigned |
Unit: | unassigned |
Preamp: | 0 |
Current location: | Bochum |
Installation information: | |
Label: | none |
Manufacturer information: | |
Wafer position: | E08 |
Break-through voltage: | 418 V |
Voltage for Gain 100 (T=+25°C): | 389.6 V |
Dark current: | 4.4 nA |
Screening Logistics: | |
Available: | Yes |
Storage Box: | 309 |
Position in Box: | 38 |
EP1 batch: | none |
EP1 batch after irradiation: | none |
Shipment: | |
Grid number: | 403 |
Position in grid: | 13 |
Arrival for irradiation: | 29. Sep 2017 |
Sent for analysis after irradiation: | 20. Oct 2017 |
Return for assembly: | none |
Irradiation: | |
Date: | 20. Oct 2017 |
Dose used: | 45 Gy |
Temperature: | 20 °C |
Position: | 3 |
Bias voltage: | 389.6 V (connected) |
Annealing: | |
Date: | none |
Temperature: | none |
Duration: | none |
Measurement results: | |
Voltage for Gain 100: | T = +20 °C: 390.0832101 V T = -25 °C: 353.9043371 V |
Voltage for Gain 150: | T = +20 °C: 398.0372883 V T = -25 °C: 361.4414919 V |
Voltage for Gain 200: | T = +20 °C: 402.4640696 V T = -25 °C: 365.6746743 V |
Gain/Voltage slope at M = 100: | T = +20 °C: 4.525783116 V-1 T = -25 °C: 4.893081677 V-1 |
Gain/Voltage slope at M = 150: | T = +20 °C: 8.906709391 V-1 T = -25 °C: 8.957544615 V-1 |
Gain/Voltage slope at M = 200: | T = +20 °C: 13.75557273 V-1 T = -25 °C: 15.73652636 V-1 |
Break-through voltage: | T = +20 °C: 418.3201862 V T = -25 °C: 381.5136628 V |
Notes: |
Version history
Time | Author | Change comment |
---|---|---|
05. Feb 2021 11:28:21 CET | jreher | Updated box and position via API. |
11. Aug 2020 17:58:44 CEST | jreher | Updated APD characteristics via API. |
21. Jul 2020 14:45:11 CEST | jreher | Updated APD location via API. |
21. Jul 2020 14:45:01 CEST | jreher | Updated box and position via API. |
19. Dec 2019 16:48:01 CET | spankonin | Updated APD batch assignment via API. |
19. Dec 2019 15:19:02 CET | jreher | Updated box and position via API. |
20. Oct 2017 14:31:54 CEST | markus.moritz | Entered shipped after irradiation date. |
20. Oct 2017 09:51:22 CEST | markus.moritz | Entered irradiation information. |
29. Sep 2017 11:33:00 CEST | markus.moritz | Entered arrival for irradiation date. |
01. Aug 2016 17:29:33 CEST | Tobias | Assigned to grid via information extracted from APD DB. |
23. Jun 2016 10:39:49 CEST | Tobias | Imported from Hamamatsu datasheet. |