RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 1409015717 - Archived data from Tue, 11. August 2020 17:25:06 CEST


This is all the information about APD 1409015717. If it is wrong, edit the data.

Subdetector specification:  
Serial: 1409015717
Type: Hamamatsu Avalanche Photo Diode
Detector: unassigned
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: none
   
Manufacturer information:  
Wafer position: D13
Break-through voltage: 425 V
Voltage for Gain 100 (T=+25°C): 396.4 V
Dark current: 6.3 nA
   
Screening Logistics:  
Available: Yes
Storage Box: 33
Position in Box: 41
EP1 batch: none
EP1 batch after irradiation: none
   
Shipment:  
Grid number: 385
Position in grid: 19
Arrival for irradiation: 29. Sep 2017
Sent for analysis after irradiation: 26. Nov 2018
Return for assembly: none
   
Irradiation:  
Date: 22. Nov 2018
Dose used: 37 Gy
Temperature: 20 °C
Position: none
Bias voltage: none     
   
Annealing:  
Date: 23. Nov 2018
Temperature: 80 °C
Duration: 48 h
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 397.0406464 V     T = -25 °C: 360.21818 V
Voltage for Gain 150: T = +20 °C: 405.0452292 V     T = -25 °C: 368.0169152 V
Voltage for Gain 200: T = +20 °C: 409.5043176 V     T = -25 °C: 372.4125853 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.348790514 V-1     T = -25 °C: 4.759335088 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 8.984133161 V-1     T = -25 °C: 9.135642195 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 15.08158795 V-1     T = -25 °C: 15.43003108 V-1
Break-through voltage: T = +20 °C: 425.3176097 V     T = -25 °C: 388.3799528 V
   
Notes:


Version history

Time Author Change comment
05. Feb 2021 10:52:09 CET jreher Updated box and position via API.
03. Feb 2021 18:18:15 CET jreher Updated box and position via API.
11. Aug 2020 17:25:06 CEST jreher Updated APD characteristics via API.
28. Nov 2019 12:22:27 CET jreher Updated APD location via API.
25. Nov 2019 17:13:30 CET jreher Updated box and position via API.
13. Jun 2019 11:29:03 CEST jreher Updated APD location via API.
12. Jun 2019 16:47:54 CEST jreher Updated APD batch assignment via API.
12. Jun 2019 16:32:49 CEST jreher Updated APD batch assignment via API.
26. Nov 2018 13:44:58 CET markus.moritz
29. Sep 2017 13:36:52 CEST markus.moritz Entered arrival for irradiation date.
01. Aug 2016 17:29:22 CEST Tobias Assigned to grid via information extracted from APD DB.
23. Jun 2016 10:32:30 CEST Tobias Imported from Hamamatsu datasheet.