Show Hamamatsu Avalanche Photo Diode 1406015496 - Archived data from Tue, 11. August 2020 17:25:33 CEST
This is all the information about APD 1406015496. If it is wrong, edit the data.
Subdetector specification: | |
---|---|
Serial: | 1406015496 |
Type: | Hamamatsu Avalanche Photo Diode |
Detector: | unassigned |
Unit: | unassigned |
Preamp: | 0 |
Current location: | Bochum |
Installation information: | |
Label: | none |
Manufacturer information: | |
Wafer position: | C11 |
Break-through voltage: | 424 V |
Voltage for Gain 100 (T=+25°C): | 395.1 V |
Dark current: | 5.2 nA |
Screening Logistics: | |
Available: | Yes |
Storage Box: | 36 |
Position in Box: | 24 |
EP1 batch: | none |
EP1 batch after irradiation: | none |
Shipment: | |
Grid number: | 375 |
Position in grid: | 7 |
Arrival for irradiation: | 29. Sep 2017 |
Sent for analysis after irradiation: | 03. Jan 2019 |
Return for assembly: | none |
Irradiation: | |
Date: | 05. Dec 2018 |
Dose used: | 37 Gy |
Temperature: | 20 °C |
Position: | none |
Bias voltage: | none |
Annealing: | |
Date: | 10. Dec 2018 |
Temperature: | 80 °C |
Duration: | 48 h |
Measurement results: | |
Voltage for Gain 100: | T = +20 °C: 395.3732742 V T = -25 °C: 358.7529169 V |
Voltage for Gain 150: | T = +20 °C: 403.2343578 V T = -25 °C: 366.4396424 V |
Voltage for Gain 200: | T = +20 °C: 407.6017627 V T = -25 °C: 370.7760955 V |
Gain/Voltage slope at M = 100: | T = +20 °C: 4.407541044 V-1 T = -25 °C: 4.533302089 V-1 |
Gain/Voltage slope at M = 150: | T = +20 °C: 9.410917656 V-1 T = -25 °C: 8.905514499 V-1 |
Gain/Voltage slope at M = 200: | T = +20 °C: 14.82740504 V-1 T = -25 °C: 14.47498289 V-1 |
Break-through voltage: | T = +20 °C: 423.4097303 V T = -25 °C: 386.6578015 V |
Notes: |
Version history
Time | Author | Change comment |
---|---|---|
05. Feb 2021 10:52:41 CET | jreher | Updated box and position via API. |
03. Feb 2021 18:18:36 CET | jreher | Updated box and position via API. |
11. Aug 2020 17:25:33 CEST | jreher | Updated APD characteristics via API. |
28. Nov 2019 12:23:00 CET | jreher | Updated APD location via API. |
25. Nov 2019 17:13:43 CET | jreher | Updated box and position via API. |
13. Jun 2019 11:29:43 CEST | jreher | Updated APD location via API. |
12. Jun 2019 16:48:06 CEST | jreher | Updated APD batch assignment via API. |
12. Jun 2019 16:33:01 CEST | jreher | Updated APD batch assignment via API. |
03. Jan 2019 12:13:20 CET | markus.moritz | |
29. Sep 2017 11:17:36 CEST | markus.moritz | Entered arrival for irradiation date. |
01. Aug 2016 17:29:15 CEST | Tobias | Assigned to grid via information extracted from APD DB. |
23. Jun 2016 10:32:23 CEST | Tobias | Imported from Hamamatsu datasheet. |