Show Hamamatsu Avalanche Photo Diode 1404015358 - Archived data from Fri, 16. April 2021 22:01:03 CEST
This is all the information about APD 1404015358. If it is wrong, edit the data.
Subdetector specification: | |
---|---|
Serial: | 1404015358 |
Type: | Hamamatsu Avalanche Photo Diode |
Detector: | unassigned |
Unit: | unassigned |
Preamp: | 0 |
Current location: | Bochum |
Installation information: | |
Label: | none |
Manufacturer information: | |
Wafer position: | B11 |
Break-through voltage: | 422 V |
Voltage for Gain 100 (T=+25°C): | 394.9 V |
Dark current: | 7.8 nA |
Screening Logistics: | |
Available: | Yes |
Storage Box: | 357 |
Position in Box: | 3 |
EP1 batch: | none |
EP1 batch after irradiation: | none |
Shipment: | |
Grid number: | 368 |
Position in grid: | 19 |
Arrival for irradiation: | 29. Sep 2017 |
Sent for analysis after irradiation: | 03. Nov 2017 |
Return for assembly: | none |
Irradiation: | |
Date: | 26. Oct 2017 |
Dose used: | 45 Gy |
Temperature: | 20 °C |
Position: | 3 |
Bias voltage: | 394.9 V (connected) |
Annealing: | |
Date: | none |
Temperature: | none |
Duration: | none |
Measurement results: | |
Voltage for Gain 100: | T = +20 °C: 394.41925 V T = -25 °C: 358.1856423 V |
Voltage for Gain 150: | T = +20 °C: 402.3988533 V T = -25 °C: 365.8962176 V |
Voltage for Gain 200: | T = +20 °C: 406.8150248 V T = -25 °C: 370.2358926 V |
Gain/Voltage slope at M = 100: | T = +20 °C: 4.376951359 V-1 T = -25 °C: 4.651655513 V-1 |
Gain/Voltage slope at M = 150: | T = +20 °C: 8.500688154 V-1 T = -25 °C: 9.150769042 V-1 |
Gain/Voltage slope at M = 200: | T = +20 °C: 14.63203445 V-1 T = -25 °C: 14.13952025 V-1 |
Break-through voltage: | T = +20 °C: 422.094014 V T = -25 °C: 385.4871291 V |
Notes: |
Version history
Time | Author | Change comment |
---|---|---|
21. Apr 2021 14:45:20 CEST | oafedulidis | Updated box and position via API. |
16. Apr 2021 22:01:03 CEST | jreher | Updated APD characteristics via API. |
05. Feb 2021 11:34:03 CET | jreher | Updated box and position via API. |
08. Oct 2020 18:15:33 CEST | jreher | Updated APD location via API. |
17. Sep 2020 14:26:28 CEST | vwerner | Updated APD batch assignment via API. |
17. Aug 2020 14:32:03 CEST | vwerner | Updated APD batch assignment via API. |
03. Nov 2017 16:13:11 CET | markus.moritz | Entered shipped after irradiation date. |
26. Oct 2017 16:36:27 CEST | markus.moritz | Entered irradiation information. |
29. Sep 2017 12:08:48 CEST | markus.moritz | Entered arrival for irradiation date. |
01. Aug 2016 17:29:11 CEST | Tobias | Assigned to grid via information extracted from APD DB. |
23. Jun 2016 10:31:21 CEST | Tobias | Imported from Hamamatsu datasheet. |