RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 1401015148 - Archived data from Tue, 11. August 2020 17:58:31 CEST


This is all the information about APD 1401015148. If it is wrong, edit the data.

Subdetector specification:  
Serial: 1401015148
Type: Hamamatsu Avalanche Photo Diode
Detector: unassigned
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: none
   
Manufacturer information:  
Wafer position: E07
Break-through voltage: 421 V
Voltage for Gain 100 (T=+25°C): 392.4 V
Dark current: 5 nA
   
Screening Logistics:  
Available: Yes
Storage Box: 308
Position in Box: 23
EP1 batch: none
EP1 batch after irradiation: none
   
Shipment:  
Grid number: 359
Position in grid: 2
Arrival for irradiation: 02. Nov 2017
Sent for analysis after irradiation: 03. Nov 2017
Return for assembly: none
   
Irradiation:  
Date: 03. Nov 2017
Dose used: 45 Gy
Temperature: 20 °C
Position: 3
Bias voltage: 392.4 V     (connected)
   
Annealing:  
Date: none
Temperature: none
Duration: none
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 393.020812 V     T = -25 °C: 356.7320897 V
Voltage for Gain 150: T = +20 °C: 400.9512583 V     T = -25 °C: 364.2685586 V
Voltage for Gain 200: T = +20 °C: 405.3624367 V     T = -25 °C: 368.5054456 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.549809154 V-1     T = -25 °C: 4.946551862 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 8.950720023 V-1     T = -25 °C: 9.036667732 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 13.82498499 V-1     T = -25 °C: 15.96613321 V-1
Break-through voltage: T = +20 °C: 421.1102604 V     T = -25 °C: 384.2604219 V
   
Notes:


Version history

Time Author Change comment
15. Nov 2022 12:56:20 CET oafedulidis Assigned to capsule by script
16. Dec 2021 13:40:39 CET oafedulidis Updated box and position via API.
05. Feb 2021 11:28:11 CET jreher Updated box and position via API.
11. Aug 2020 17:58:31 CEST jreher Updated APD characteristics via API.
21. Jul 2020 14:44:55 CEST jreher Updated APD location via API.
21. Jul 2020 14:44:45 CEST jreher Updated box and position via API.
19. Dec 2019 16:36:31 CET spankonin Updated APD batch assignment via API.
19. Dec 2019 15:18:53 CET jreher Updated box and position via API.
03. Nov 2017 16:15:16 CET markus.moritz Entered shipped after irradiation date.
03. Nov 2017 16:14:32 CET markus.moritz Entered irradiation information.
02. Nov 2017 15:18:22 CET markus.moritz Entered arrival for irradiation date.
01. Aug 2016 17:29:04 CEST Tobias Assigned to grid via information extracted from APD DB.
23. Jun 2016 10:31:16 CEST Tobias Imported from Hamamatsu datasheet.