RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 0912009864 - Archived data from Fri, 16. April 2021 22:00:02 CEST


This is all the information about APD 0912009864. If it is wrong, edit the data.

Subdetector specification:  
Serial: 0912009864
Type: Hamamatsu Avalanche Photo Diode
Detector: unassigned
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: none
   
Manufacturer information:  
Wafer position: F06
Break-through voltage: 411 V
Voltage for Gain 100 (T=+25°C): 382.8 V
Dark current: 3.7 nA
   
Screening Logistics:  
Available: Yes
Storage Box: 347
Position in Box: 12
EP1 batch: none
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: none
Sent for analysis after irradiation: none
Return for assembly: none
   
Irradiation:  
Date: none
Dose used: none
Temperature: none
Position: none
Bias voltage: none     
   
Annealing:  
Date: none
Temperature: none
Duration: none
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 382.6490754 V     T = -25 °C: 347.0416983 V
Voltage for Gain 150: T = +20 °C: 390.5888998 V     T = -25 °C: 354.4754461 V
Voltage for Gain 200: T = +20 °C: 394.9954348 V     T = -25 °C: 358.6498133 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.358173537 V-1     T = -25 °C: 4.899582829 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 9.23631145 V-1     T = -25 °C: 8.943817652 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 14.32135624 V-1     T = -25 °C: 15.7010483 V-1
Break-through voltage: T = +20 °C: 410.7517876 V     T = -25 °C: 374.2629486 V
   
Notes:


Version history

Time Author Change comment
27. Oct 2021 14:11:21 CEST oafedulidis Assigned to capsule by script
16. Apr 2021 22:00:02 CEST jreher Updated APD characteristics via API.
15. Apr 2021 17:01:26 CEST jreher Updated box and position via API.
05. Feb 2021 11:32:47 CET jreher Updated box and position via API.
17. Aug 2020 13:44:43 CEST vwerner Updated APD batch assignment via API.
11. Aug 2020 17:54:31 CEST jreher Updated APD location via API.
23. Jun 2016 10:26:39 CEST Tobias Imported from Hamamatsu datasheet.