RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 1314014571 - Archived data from Tue, 11. August 2020 18:09:22 CEST


This is all the information about APD 1314014571. If it is wrong, edit the data.

Subdetector specification:  
Serial: 1314014571
Type: Hamamatsu Avalanche Photo Diode
Detector: unassigned
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: none
   
Manufacturer information:  
Wafer position: B12
Break-through voltage: 451 V
Voltage for Gain 100 (T=+25°C): 423.4 V
Dark current: 10.3 nA
   
Screening Logistics:  
Available: Yes
Storage Box: 326
Position in Box: 21
EP1 batch: none
EP1 batch after irradiation: none
   
Shipment:  
Grid number: 334
Position in grid: 0
Arrival for irradiation: none
Sent for analysis after irradiation: none
Return for assembly: none
   
Irradiation:  
Date: none
Dose used: none
Temperature: none
Position: none
Bias voltage: none     
   
Annealing:  
Date: none
Temperature: none
Duration: none
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 411.6876521 V     T = -25 °C: 387.2720376 V
Voltage for Gain 150: T = +20 °C: 422.347491 V     T = -25 °C: 395.1032648 V
Voltage for Gain 200: T = +20 °C: 428.4480813 V     T = -25 °C: 399.1364995 V
Gain/Voltage slope at M = 100: T = +20 °C: 3.342151523 V-1     T = -25 °C: 4.502381079 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 6.412547414 V-1     T = -25 °C: 8.962326452 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 10.13474079 V-1     T = -25 °C: 20.79968633 V-1
Break-through voltage: T = +20 °C: 440.0059103 V     T = -25 °C: 414.497541 V
   
Notes:


Version history

Time Author Change comment
07. Feb 2022 16:17:21 CET jreher Updated APD characteristics via API.
10. Nov 2021 14:35:19 CET oafedulidis Updated APD location via API.
10. Nov 2021 14:35:05 CET oafedulidis Updated APD batch assignment via API.
05. Feb 2021 11:30:10 CET jreher Updated box and position via API.
07. Oct 2020 15:45:17 CEST jreher Updated APD location via API.
11. Aug 2020 18:09:22 CEST jreher Updated APD characteristics via API.
21. Jul 2020 14:46:05 CEST jreher Updated APD location via API.
21. Jul 2020 14:45:56 CEST jreher Updated box and position via API.
09. Jul 2020 15:02:38 CEST jreher Updated APD batch assignment via API.
27. Oct 2017 10:23:25 CEST Tobias Resetted arrival for irradiation date as APDs are returned to GSI un-irradiated.
29. Sep 2017 09:53:30 CEST markus.moritz Entered arrival for irradiation date.
01. Aug 2016 17:28:47 CEST Tobias Assigned to grid via information extracted from APD DB.
22. Jun 2016 15:42:37 CEST Tobias Imported from Hamamatsu datasheet.