RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 1311014378 - Archived data from Thu, 24. February 2022 08:35:29 CET


This is all the information about APD 1311014378. If it is wrong, edit the data.

Subdetector specification:  
Serial: 1311014378
Type: Hamamatsu Avalanche Photo Diode
Detector: unassigned
Unit: unassigned
Preamp: 0
Current location: In transit to Mainz
   
Installation information:  
Label: none
   
Manufacturer information:  
Wafer position: C15
Break-through voltage: 438 V
Voltage for Gain 100 (T=+25°C): 413.6 V
Dark current: 12.9 nA
   
Screening Logistics:  
Available: Yes
Storage Box: 323
Position in Box: 17
EP1 batch: none
EP1 batch after irradiation: none
   
Shipment:  
Grid number: 326
Position in grid: 0
Arrival for irradiation: none
Sent for analysis after irradiation: none
Return for assembly: none
   
Irradiation:  
Date: none
Dose used: none
Temperature: none
Position: none
Bias voltage: none     
   
Annealing:  
Date: none
Temperature: none
Duration: none
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 413.2481683 V     T = -25 °C: 376.4629812 V
Voltage for Gain 150: T = +20 °C: 421.1200053 V     T = -25 °C: 384.3162957 V
Voltage for Gain 200: T = +20 °C: 425.4747893 V     T = -25 °C: 388.6732017 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.64689779 V-1     T = -25 °C: 4.474848689 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 8.871385623 V-1     T = -25 °C: 8.21139356 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 15.61828023 V-1     T = -25 °C: 15.51941129 V-1
Break-through voltage: T = +20 °C: 431.0030378 V     T = -25 °C: 402.0275312 V
   
Notes:


Version history

Time Author Change comment
28. Jul 2022 13:18:06 CEST oafedulidis Updated box and position via API.
28. Jul 2022 13:15:21 CEST oafedulidis Updated APD location via API.
08. Mar 2022 14:29:52 CET oafedulidis Updated APD characteristics via API.
08. Mar 2022 13:54:16 CET oafedulidis Updated APD batch assignment via API.
28. Feb 2022 15:57:49 CET oafedulidis Updated APD batch assignment via API.
24. Feb 2022 08:35:50 CET oafedulidis Updated APD location via API.
24. Feb 2022 08:35:29 CET oafedulidis Updated APD batch assignment via API.
16. Apr 2021 21:58:44 CEST jreher Updated APD characteristics via API.
05. Feb 2021 11:29:46 CET jreher Updated box and position via API.
07. Oct 2020 15:45:11 CEST jreher Updated APD location via API.
11. Aug 2020 17:53:14 CEST jreher Updated APD location via API.
21. Jul 2020 16:26:38 CEST jreher Updated APD batch assignment via API.
03. Nov 2017 11:27:24 CET Tobias Resetted arrival for irradiation date because APD was returned to GSI un-irradiated.
29. Sep 2017 12:29:21 CEST markus.moritz Entered arrival for irradiation date.
01. Aug 2016 17:28:41 CEST Tobias Assigned to grid via information extracted from APD DB.
22. Jun 2016 15:42:33 CEST Tobias Imported from Hamamatsu datasheet.