Show Hamamatsu Avalanche Photo Diode 1306013951 - Archived data from Tue, 11. August 2020 17:40:23 CEST
This is all the information about APD 1306013951. If it is wrong, edit the data.
Subdetector specification: | |
---|---|
Serial: | 1306013951 |
Type: | Hamamatsu Avalanche Photo Diode |
Detector: | unassigned |
Unit: | unassigned |
Preamp: | 0 |
Current location: | Bochum |
Installation information: | |
Label: | none |
Manufacturer information: | |
Wafer position: | G07 |
Break-through voltage: | 433.7 V |
Voltage for Gain 100 (T=+25°C): | 405.8 V |
Dark current: | 8.38 nA |
Screening Logistics: | |
Available: | Yes |
Storage Box: | 290 |
Position in Box: | 31 |
EP1 batch: | none |
EP1 batch after irradiation: | none |
Shipment: | |
Grid number: | 306 |
Position in grid: | 2 |
Arrival for irradiation: | 27. Oct 2016 |
Sent for analysis after irradiation: | 02. Nov 2016 |
Return for assembly: | 20. Dec 2016 |
Irradiation: | |
Date: | 31. Oct 2016 |
Dose used: | 30 Gy |
Temperature: | 20 °C |
Position: | 1 |
Bias voltage: | 350 V (connected) |
Annealing: | |
Date: | none |
Temperature: | none |
Duration: | none |
Measurement results: | |
Voltage for Gain 100: | T = +20 °C: 405.9731594 V T = -25 °C: 369.44317 V |
Voltage for Gain 150: | T = +20 °C: 413.8906723 V T = -25 °C: 377.2911136 V |
Voltage for Gain 200: | T = +20 °C: 418.2842442 V T = -25 °C: 381.6539388 V |
Gain/Voltage slope at M = 100: | T = +20 °C: 4.574465398 V-1 T = -25 °C: 4.46033902 V-1 |
Gain/Voltage slope at M = 150: | T = +20 °C: 9.029764543 V-1 T = -25 °C: 8.738968094 V-1 |
Gain/Voltage slope at M = 200: | T = +20 °C: 14.05879901 V-1 T = -25 °C: 15.24343826 V-1 |
Break-through voltage: | T = +20 °C: 431.0411694 V T = -25 °C: 397.2822865 V |
Notes: |
Version history
Time | Author | Change comment |
---|---|---|
05. Feb 2021 11:25:43 CET | jreher | Updated box and position via API. |
11. Aug 2020 17:40:23 CEST | jreher | Updated APD characteristics via API. |
21. Jul 2020 14:40:41 CEST | jreher | Updated box and position via API. |
27. Nov 2019 20:18:50 CET | jreher | Updated APD location via API. |
25. Nov 2019 17:31:29 CET | jreher | Updated box and position via API. |
22. Oct 2019 14:32:00 CEST | spankonin | Updated APD location via API. |
15. Oct 2019 11:02:53 CEST | jgrochowski | Updated APD batch assignment via API. |
20. Dec 2016 10:39:41 CET | markus.moritz | Entered return for assembly date. |
02. Nov 2016 14:30:30 CET | markus.moritz | Entered shipped after irradiation date. |
02. Nov 2016 14:19:18 CET | markus.moritz | Entered irradiation information. |
27. Oct 2016 12:06:35 CEST | markus.moritz | Entered arrival for irradiation date. |
01. Aug 2016 17:28:27 CEST | Tobias | Assigned to grid via information extracted from APD DB. |
22. Jun 2016 15:41:01 CEST | Tobias | Imported from Hamamatsu datasheet. |