RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 1302013685 - Archived data from Tue, 11. August 2020 18:03:55 CEST


This is all the information about APD 1302013685. If it is wrong, edit the data.

Subdetector specification:  
Serial: 1302013685
Type: Hamamatsu Avalanche Photo Diode
Detector: unassigned
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: none
   
Manufacturer information:  
Wafer position: D09
Break-through voltage: 440 V
Voltage for Gain 100 (T=+25°C): 412 V
Dark current: 6.6 nA
   
Screening Logistics:  
Available: Yes
Storage Box: 284
Position in Box: 37
EP1 batch: 202
EP1 batch after irradiation: none
   
Shipment:  
Grid number: 293
Position in grid: 13
Arrival for irradiation: 29. Sep 2017
Sent for analysis after irradiation: none
Return for assembly: none
   
Irradiation:  
Date: none
Dose used: none
Temperature: none
Position: none
Bias voltage: none     
   
Annealing:  
Date: none
Temperature: none
Duration: none
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 412.721175 V     T = -25 °C: 375.7405335 V
Voltage for Gain 150: T = +20 °C: 420.7081661 V     T = -25 °C: 383.6781608 V
Voltage for Gain 200: T = +20 °C: 425.1293203 V     T = -25 °C: 388.0735464 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.600529875 V-1     T = -25 °C: 4.637190386 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 9.059617185 V-1     T = -25 °C: 9.140659057 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 14.10477921 V-1     T = -25 °C: 14.27953447 V-1
Break-through voltage: T = +20 °C: 439.7867935 V     T = -25 °C: 403.342273 V
   
Notes:


Version history

Time Author Change comment
05. Feb 2021 11:24:59 CET jreher Updated box and position via API.
07. Oct 2020 15:44:40 CEST jreher Updated APD location via API.
11. Aug 2020 18:03:55 CEST jreher Updated APD characteristics via API.
21. Jul 2020 14:39:39 CEST jreher Updated APD location via API.
21. Jul 2020 14:39:30 CEST jreher Updated box and position via API.
10. Feb 2020 18:58:27 CET spankonin Updated APD batch assignment via API.
29. Nov 2019 11:50:12 CET jreher Updated APD location via API.
27. Nov 2019 20:17:49 CET jreher Updated APD location via API.
25. Nov 2019 17:31:01 CET jreher Updated box and position via API.
22. Oct 2019 13:10:00 CEST spankonin Updated APD location via API.
09. Oct 2019 11:41:28 CEST jgrochowski Updated APD batch assignment via API.
29. Sep 2017 12:03:34 CEST markus.moritz Entered arrival for irradiation date.
01. Aug 2016 17:28:19 CEST Tobias Assigned to grid via information extracted from APD DB.
22. Jun 2016 15:39:45 CEST Tobias Imported from Hamamatsu datasheet.