Show Hamamatsu Avalanche Photo Diode 1204012959 - Archived data from Tue, 11. August 2020 18:10:12 CEST
This is all the information about APD 1204012959. If it is wrong, edit the data.
Subdetector specification: | |
---|---|
Serial: | 1204012959 |
Type: | Hamamatsu Avalanche Photo Diode |
Detector: | unassigned |
Unit: | unassigned |
Preamp: | 0 |
Current location: | Bochum |
Installation information: | |
Label: | none |
Manufacturer information: | |
Wafer position: | G08 |
Break-through voltage: | 435 V |
Voltage for Gain 100 (T=+25°C): | 405.8 V |
Dark current: | 9.9 nA |
Screening Logistics: | |
Available: | Yes |
Storage Box: | 275 |
Position in Box: | 45 |
EP1 batch: | none |
EP1 batch after irradiation: | none |
Shipment: | |
Grid number: | 262 |
Position in grid: | 0 |
Arrival for irradiation: | 29. Sep 2017 |
Sent for analysis after irradiation: | none |
Return for assembly: | none |
Irradiation: | |
Date: | none |
Dose used: | none |
Temperature: | none |
Position: | none |
Bias voltage: | none |
Annealing: | |
Date: | none |
Temperature: | none |
Duration: | none |
Measurement results: | |
Voltage for Gain 100: | T = +20 °C: 406.3608558 V T = -25 °C: 369.7382817 V |
Voltage for Gain 150: | T = +20 °C: 414.3112572 V T = -25 °C: 377.5961101 V |
Voltage for Gain 200: | T = +20 °C: 418.7110938 V T = -25 °C: 381.9418588 V |
Gain/Voltage slope at M = 100: | T = +20 °C: 4.377889041 V-1 T = -25 °C: 4.677505202 V-1 |
Gain/Voltage slope at M = 150: | T = +20 °C: 8.520021883 V-1 T = -25 °C: 9.355390532 V-1 |
Gain/Voltage slope at M = 200: | T = +20 °C: 15.11060419 V-1 T = -25 °C: 14.63648999 V-1 |
Break-through voltage: | T = +20 °C: 426.6311406 V T = -25 °C: 398.2853676 V |
Notes: |
Version history
Time | Author | Change comment |
---|---|---|
05. Feb 2021 11:23:50 CET | jreher | Updated box and position via API. |
11. Aug 2020 18:10:12 CEST | jreher | Updated APD characteristics via API. |
21. Jul 2020 14:37:27 CEST | jreher | Updated APD location via API. |
21. Jul 2020 14:37:17 CEST | jreher | Updated box and position via API. |
29. Jun 2020 16:14:06 CEST | jreher | Updated APD batch assignment via API. |
29. Sep 2017 11:22:27 CEST | markus.moritz | Entered arrival for irradiation date. |
01. Aug 2016 17:27:58 CEST | Tobias | Assigned to grid via information extracted from APD DB. |
22. Jun 2016 15:37:53 CEST | Tobias | Imported from Hamamatsu datasheet. |