RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 0905009355 - Archived data from Fri, 16. April 2021 22:00:52 CEST


This is all the information about APD 0905009355. If it is wrong, edit the data.

Subdetector specification:  
Serial: 0905009355
Type: Hamamatsu Avalanche Photo Diode
Detector: unassigned
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: none
   
Manufacturer information:  
Wafer position: F07
Break-through voltage: 402 V
Voltage for Gain 100 (T=+25°C): 374.3 V
Dark current: 5.3 nA
   
Screening Logistics:  
Available: Yes
Storage Box: 355
Position in Box: 7
EP1 batch: none
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: none
Sent for analysis after irradiation: none
Return for assembly: none
   
Irradiation:  
Date: none
Dose used: none
Temperature: none
Position: none
Bias voltage: none     
   
Annealing:  
Date: none
Temperature: none
Duration: none
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 374.135747 V     T = -25 °C: 339.4168375 V
Voltage for Gain 150: T = +20 °C: 381.9210093 V     T = -25 °C: 346.6040897 V
Voltage for Gain 200: T = +20 °C: 386.270036 V     T = -25 °C: 350.6273384 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.596210204 V-1     T = -25 °C: 4.913459422 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 9.053434969 V-1     T = -25 °C: 10.0842106 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 14.04355401 V-1     T = -25 °C: 16.29641019 V-1
Break-through voltage: T = +20 °C: 401.1049891 V     T = -25 °C: 365.2764104 V
   
Notes:


Version history

Time Author Change comment
21. Apr 2021 14:46:42 CEST oafedulidis Updated box and position via API.
16. Apr 2021 22:00:52 CEST jreher Updated APD characteristics via API.
05. Feb 2021 11:33:49 CET jreher Updated box and position via API.
08. Oct 2020 18:15:10 CEST jreher Updated APD location via API.
17. Aug 2020 14:25:44 CEST vwerner Updated APD batch assignment via API.
22. Jun 2016 15:31:46 CEST Tobias Imported from Hamamatsu datasheet.