Show Hamamatsu Avalanche Photo Diode 0902009175 - Archived data from Tue, 11. August 2020 17:56:17 CEST
This is all the information about APD 0902009175. If it is wrong, edit the data.
Subdetector specification: | |
---|---|
Serial: | 0902009175 |
Type: | Hamamatsu Avalanche Photo Diode |
Detector: | unassigned |
Unit: | unassigned |
Preamp: | 0 |
Current location: | Bochum |
Installation information: | |
Label: | none |
Manufacturer information: | |
Wafer position: | E12 |
Break-through voltage: | 408 V |
Voltage for Gain 100 (T=+25°C): | 379.6 V |
Dark current: | 8.4 nA |
Screening Logistics: | |
Available: | Yes |
Storage Box: | 295 |
Position in Box: | 2 |
EP1 batch: | none |
EP1 batch after irradiation: | none |
Shipment: | |
Grid number: | 183 |
Position in grid: | 10 |
Arrival for irradiation: | 30. Oct 2017 |
Sent for analysis after irradiation: | 17. Nov 2017 |
Return for assembly: | none |
Irradiation: | |
Date: | 16. Nov 2017 |
Dose used: | 45 Gy |
Temperature: | 20 °C |
Position: | 1 |
Bias voltage: | 379.6 V (connected) |
Annealing: | |
Date: | none |
Temperature: | none |
Duration: | none |
Measurement results: | |
Voltage for Gain 100: | T = +20 °C: 379.5945271 V T = -25 °C: 344.3833369 V |
Voltage for Gain 150: | T = +20 °C: 387.5736715 V T = -25 °C: 351.860155 V |
Voltage for Gain 200: | T = +20 °C: 391.9934373 V T = -25 °C: 355.9570437 V |
Gain/Voltage slope at M = 100: | T = +20 °C: 4.379179953 V-1 T = -25 °C: 5.382091732 V-1 |
Gain/Voltage slope at M = 150: | T = +20 °C: 8.380805109 V-1 T = -25 °C: 9.900146416 V-1 |
Gain/Voltage slope at M = 200: | T = +20 °C: 14.35862892 V-1 T = -25 °C: 14.67621665 V-1 |
Break-through voltage: | T = +20 °C: 407.7297344 V T = -25 °C: 371.3521102 V |
Notes: | Wrong wafer number in serial number, correct wafer number is 3. |
Version history
Time | Author | Change comment |
---|---|---|
05. Feb 2021 11:26:18 CET | jreher | Updated box and position via API. |
11. Aug 2020 17:56:17 CEST | jreher | Updated APD characteristics via API. |
21. Jul 2020 14:41:32 CEST | jreher | Updated box and position via API. |
19. Dec 2019 15:44:46 CET | spankonin | Updated APD batch assignment via API. |
19. Dec 2019 15:17:17 CET | jreher | Updated APD location via API. |
19. Dec 2019 15:17:10 CET | jreher | Updated box and position via API. |
17. Nov 2017 09:39:16 CET | markus.moritz | Entered shipped after irradiation date. |
17. Nov 2017 09:38:13 CET | markus.moritz | Entered irradiation information. |
09. Nov 2017 15:11:05 CET | markus.moritz | Entered arrival for irradiation date. |
02. Nov 2017 15:16:53 CET | Tobias | Assigned to grid via information extracted from APD DB. |
23. Jun 2016 10:03:46 CEST | Tobias | Added note about wrong wafer number. |
22. Jun 2016 15:31:42 CEST | Tobias | Imported from Hamamatsu datasheet. |