Show Hamamatsu Avalanche Photo Diode 0810007969 - Archived data from Fri, 05. February 2021 11:26:28 CET
This is all the information about APD 0810007969. If it is wrong, edit the data.
Subdetector specification: | |
---|---|
Serial: | 0810007969 |
Type: | Hamamatsu Avalanche Photo Diode |
Detector: | unassigned |
Unit: | unassigned |
Preamp: | 0 |
Current location: | Bochum |
Installation information: | |
Label: | none |
Manufacturer information: | |
Wafer position: | E12 |
Break-through voltage: | 415 V |
Voltage for Gain 100 (T=+25°C): | 387.4 V |
Dark current: | 4.3 nA |
Screening Logistics: | |
Available: | Yes |
Storage Box: | 296 |
Position in Box: | 13 |
EP1 batch: | none |
EP1 batch after irradiation: | none |
Shipment: | |
Grid number: | 103 |
Position in grid: | 4 |
Arrival for irradiation: | 03. Aug 2017 |
Sent for analysis after irradiation: | 23. Aug 2017 |
Return for assembly: | none |
Irradiation: | |
Date: | 10. Aug 2017 |
Dose used: | 45 Gy |
Temperature: | 20 °C |
Position: | 3 |
Bias voltage: | 387.4 V (connected) |
Annealing: | |
Date: | none |
Temperature: | none |
Duration: | none |
Measurement results: | |
Voltage for Gain 100: | T = +20 °C: 387.5389253 V T = -25 °C: 351.4888939 V |
Voltage for Gain 150: | T = +20 °C: 395.3958488 V T = -25 °C: 358.9259127 V |
Voltage for Gain 200: | T = +20 °C: 399.7456075 V T = -25 °C: 363.0904829 V |
Gain/Voltage slope at M = 100: | T = +20 °C: 4.431407085 V-1 T = -25 °C: 4.697182437 V-1 |
Gain/Voltage slope at M = 150: | T = +20 °C: 8.671344069 V-1 T = -25 °C: 9.510074521 V-1 |
Gain/Voltage slope at M = 200: | T = +20 °C: 15.06597286 V-1 T = -25 °C: 15.01363505 V-1 |
Break-through voltage: | T = +20 °C: 414.6274732 V T = -25 °C: 378.1098206 V |
Notes: |
Version history
Time | Author | Change comment |
---|---|---|
21. Apr 2021 14:48:04 CEST | oafedulidis | Updated box and position via API. |
05. Feb 2021 11:26:28 CET | jreher | Updated box and position via API. |
11. Aug 2020 17:56:30 CEST | jreher | Updated APD characteristics via API. |
21. Jul 2020 14:41:56 CEST | jreher | Updated APD location via API. |
21. Jul 2020 14:41:47 CEST | jreher | Updated box and position via API. |
19. Dec 2019 15:47:37 CET | spankonin | Updated APD batch assignment via API. |
19. Dec 2019 15:17:20 CET | jreher | Updated box and position via API. |
23. Aug 2017 15:55:38 CEST | markus.moritz | Entered shipped after irradiation date. |
10. Aug 2017 16:43:53 CEST | hans-georg.zaunick | Entered irradiation information. |
07. Aug 2017 17:01:09 CEST | hans-georg.zaunick | Entered arrival for irradiation date. |
04. Aug 2017 03:15:28 CEST | Tobias | Assigned to grid via information extracted from APD DB. |
22. Jun 2016 15:26:14 CEST | Tobias | Imported from Hamamatsu datasheet. |