RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 3314036309 - Archived data from Tue, 11. August 2020 18:02:39 CEST


This is all the information about APD 3314036309. If it is wrong, edit the data.

Subdetector specification:  
Serial: 3314036309
Type: Hamamatsu Avalanche Photo Diode
Detector: unassigned
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: none
   
Manufacturer information:  
Wafer position: E13
Break-through voltage: 412 V
Voltage for Gain 100 (T=+25°C): 384 V
Dark current: 22.1 nA
   
Screening Logistics:  
Available: Yes
Storage Box: 154
Position in Box: 31
EP1 batch: 182
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: none
Sent for analysis after irradiation: none
Return for assembly: none
   
Irradiation:  
Date: none
Dose used: none
Temperature: none
Position: none
Bias voltage: none     
   
Annealing:  
Date: none
Temperature: none
Duration: none
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 384.0371509 V     T = -25 °C: 349.2043585 V
Voltage for Gain 150: T = +20 °C: 391.781895 V     T = -25 °C: 356.4382992 V
Voltage for Gain 200: T = +20 °C: 396.096675 V     T = -25 °C: 360.5039755 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.69159192 V-1     T = -25 °C: 4.953188645 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 9.269503095 V-1     T = -25 °C: 9.266108168 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 14.50915392 V-1     T = -25 °C: 16.63436386 V-1
Break-through voltage: T = +20 °C: 411.5129746 V     T = -25 °C: 375.5262502 V
   
Notes:


Version history

Time Author Change comment
05. Feb 2021 11:08:08 CET jreher Updated box and position via API.
11. Aug 2020 18:02:39 CEST jreher Updated APD characteristics via API.
11. Aug 2020 17:52:19 CEST jreher Updated APD location via API.
03. Feb 2020 14:22:28 CET spankonin Updated APD batch assignment via API.
29. Nov 2019 11:49:28 CET jreher Updated APD location via API.
27. Nov 2019 19:56:05 CET jreher Updated APD location via API.
25. Nov 2019 17:23:06 CET jreher Updated box and position via API.
01. Oct 2019 11:32:16 CEST spankonin Updated APD batch assignment via API.
05. Jul 2019 13:56:42 CEST jreher Updated APD location via API.
03. Jan 2019 00:15:25 CET Tobias Corrected dark current unit.
19. Sep 2018 14:05:34 CEST Tobias Imported from Hamamatsu datasheet.