RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 3214035388 - Archived data from Tue, 11. August 2020 17:55:04 CEST


This is all the information about APD 3214035388. If it is wrong, edit the data.

Subdetector specification:  
Serial: 3214035388
Type: Hamamatsu Avalanche Photo Diode
Detector: unassigned
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: none
   
Manufacturer information:  
Wafer position: F06
Break-through voltage: 409 V
Voltage for Gain 100 (T=+25°C): 380.8 V
Dark current: 24.1 nA
   
Screening Logistics:  
Available: Yes
Storage Box: 142
Position in Box: 2
EP1 batch: 164
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: none
Sent for analysis after irradiation: none
Return for assembly: none
   
Irradiation:  
Date: none
Dose used: none
Temperature: none
Position: none
Bias voltage: none     
   
Annealing:  
Date: none
Temperature: none
Duration: none
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 380.495491 V     T = -25 °C: 345.4033267 V
Voltage for Gain 150: T = +20 °C: 388.3459632 V     T = -25 °C: 352.7077294 V
Voltage for Gain 200: T = +20 °C: 392.7115537 V     T = -25 °C: 356.8042168 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.456087348 V-1     T = -25 °C: 4.817577357 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 8.657402774 V-1     T = -25 °C: 9.90089066 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 15.04311598 V-1     T = -25 °C: 15.81147135 V-1
Break-through voltage: T = +20 °C: 408.3522538 V     T = -25 °C: 372.0231973 V
   
Notes:


Version history

Time Author Change comment
17. Jun 2021 19:22:30 CEST oafedulidis Assigned to capsule by script
16. Feb 2021 12:27:21 CET jreher Updated box and position via API.
05. Feb 2021 11:06:37 CET jreher Updated box and position via API.
11. Aug 2020 17:55:04 CEST jreher Updated APD characteristics via API.
11. Aug 2020 17:51:27 CEST jreher Updated APD location via API.
12. Dec 2019 14:41:09 CET spankonin Updated APD batch assignment via API.
29. Nov 2019 11:48:43 CET jreher Updated APD location via API.
27. Nov 2019 19:53:54 CET jreher Updated APD location via API.
25. Nov 2019 17:22:08 CET jreher Updated box and position via API.
20. Sep 2019 18:49:47 CEST spankonin Updated APD batch assignment via API.
05. Jul 2019 13:55:12 CEST jreher Updated APD location via API.
03. Jan 2019 00:15:05 CET Tobias Corrected dark current unit.
14. Aug 2018 16:04:24 CEST Tobias Imported from Hamamatsu datasheet.