RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 3213035350 - Archived data from Tue, 11. August 2020 17:54:59 CEST


This is all the information about APD 3213035350. If it is wrong, edit the data.

Subdetector specification:  
Serial: 3213035350
Type: Hamamatsu Avalanche Photo Diode
Detector: unassigned
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: none
   
Manufacturer information:  
Wafer position: G07
Break-through voltage: 419 V
Voltage for Gain 100 (T=+25°C): 392.2 V
Dark current: 28.4 nA
   
Screening Logistics:  
Available: Yes
Storage Box: 141
Position in Box: 27
EP1 batch: 163
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: none
Sent for analysis after irradiation: none
Return for assembly: none
   
Irradiation:  
Date: none
Dose used: none
Temperature: none
Position: none
Bias voltage: none     
   
Annealing:  
Date: none
Temperature: none
Duration: none
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 390.9168012 V     T = -25 °C: 354.6272775 V
Voltage for Gain 150: T = +20 °C: 398.7469291 V     T = -25 °C: 362.1599964 V
Voltage for Gain 200: T = +20 °C: 403.0710357 V     T = -25 °C: 366.3899149 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.64948235 V-1     T = -25 °C: 4.618554208 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 9.267561494 V-1     T = -25 °C: 9.169792982 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 14.59264259 V-1     T = -25 °C: 14.39053622 V-1
Break-through voltage: T = +20 °C: 417.3284239 V     T = -25 °C: 380.7596387 V
   
Notes:


Version history

Time Author Change comment
05. Feb 2021 11:06:32 CET jreher Updated box and position via API.
11. Aug 2020 17:54:59 CEST jreher Updated APD characteristics via API.
11. Aug 2020 17:51:24 CEST jreher Updated APD location via API.
12. Dec 2019 14:41:07 CET spankonin Updated APD batch assignment via API.
29. Nov 2019 11:48:41 CET jreher Updated APD location via API.
27. Nov 2019 19:53:45 CET jreher Updated APD location via API.
25. Nov 2019 17:22:05 CET jreher Updated box and position via API.
20. Sep 2019 18:49:15 CEST spankonin Updated APD batch assignment via API.
05. Jul 2019 13:55:06 CEST jreher Updated APD location via API.
03. Jan 2019 00:15:04 CET Tobias Corrected dark current unit.
14. Aug 2018 16:04:23 CEST Tobias Imported from Hamamatsu datasheet.