RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 3116034739 - Archived data from Tue, 11. August 2020 17:54:06 CEST


This is all the information about APD 3116034739. If it is wrong, edit the data.

Subdetector specification:  
Serial: 3116034739
Type: Hamamatsu Avalanche Photo Diode
Detector: unassigned
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: none
   
Manufacturer information:  
Wafer position: E07
Break-through voltage: 402 V
Voltage for Gain 100 (T=+25°C): 374 V
Dark current: 12.2 nA
   
Screening Logistics:  
Available: Yes
Storage Box: 132
Position in Box: 18
EP1 batch: 149
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: none
Sent for analysis after irradiation: none
Return for assembly: none
   
Irradiation:  
Date: none
Dose used: none
Temperature: none
Position: none
Bias voltage: none     
   
Annealing:  
Date: none
Temperature: none
Duration: none
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 373.6602084 V     T = -25 °C: 339.3183846 V
Voltage for Gain 150: T = +20 °C: 381.436863 V     T = -25 °C: 346.4999111 V
Voltage for Gain 200: T = +20 °C: 385.7903332 V     T = -25 °C: 350.5371598 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.764800948 V-1     T = -25 °C: 4.939903273 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 8.567548532 V-1     T = -25 °C: 10.14802635 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 14.63918444 V-1     T = -25 °C: 16.26919157 V-1
Break-through voltage: T = +20 °C: 401.5972204 V     T = -25 °C: 365.6420173 V
   
Notes:


Version history

Time Author Change comment
05. Feb 2021 11:05:23 CET jreher Updated box and position via API.
11. Aug 2020 17:54:06 CEST jreher Updated APD characteristics via API.
11. Aug 2020 17:50:44 CEST jreher Updated APD location via API.
09. Dec 2019 18:15:30 CET spankonin Updated APD batch assignment via API.
29. Nov 2019 11:48:09 CET jreher Updated APD location via API.
27. Nov 2019 19:52:06 CET jreher Updated APD location via API.
25. Nov 2019 17:21:24 CET jreher Updated box and position via API.
17. Sep 2019 12:08:20 CEST jgrochowski Updated APD batch assignment via API.
05. Jul 2019 13:53:57 CEST jreher Updated APD location via API.
03. Jan 2019 00:14:50 CET Tobias Corrected dark current unit.
14. Aug 2018 16:04:15 CEST Tobias Imported from Hamamatsu datasheet.