RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 3009033761 - Archived data from Tue, 11. August 2020 18:07:14 CEST


This is all the information about APD 3009033761. If it is wrong, edit the data.

Subdetector specification:  
Serial: 3009033761
Type: Hamamatsu Avalanche Photo Diode
Detector: unassigned
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: none
   
Manufacturer information:  
Wafer position: E12
Break-through voltage: 415 V
Voltage for Gain 100 (T=+25°C): 387.4 V
Dark current: 19.9 nA
   
Screening Logistics:  
Available: Yes
Storage Box: 118
Position in Box: 44
EP1 batch: 127
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: none
Sent for analysis after irradiation: none
Return for assembly: none
   
Irradiation:  
Date: none
Dose used: none
Temperature: none
Position: none
Bias voltage: none     
   
Annealing:  
Date: none
Temperature: none
Duration: none
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 387.411463 V     T = -25 °C: 351.7501866 V
Voltage for Gain 150: T = +20 °C: 395.2248712 V     T = -25 °C: 359.2153034 V
Voltage for Gain 200: T = +20 °C: 399.5610071 V     T = -25 °C: 363.4102158 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.458801272 V-1     T = -25 °C: 4.951466823 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 8.762022463 V-1     T = -25 °C: 9.070437227 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 15.25733436 V-1     T = -25 °C: 16.03929819 V-1
Break-through voltage: T = +20 °C: 414.5874141 V     T = -25 °C: 378.5286884 V
   
Notes:


Version history

Time Author Change comment
05. Feb 2021 11:03:39 CET jreher Updated box and position via API.
11. Aug 2020 18:07:14 CEST jreher Updated APD characteristics via API.
11. Aug 2020 17:49:43 CEST jreher Updated APD location via API.
02. Mar 2020 16:21:38 CET vwerner Updated APD batch assignment via API.
29. Nov 2019 11:47:18 CET jreher Updated APD location via API.
27. Nov 2019 19:49:34 CET jreher Updated APD location via API.
25. Nov 2019 17:20:19 CET jreher Updated box and position via API.
30. Aug 2019 11:27:48 CEST jgrochowski Updated APD batch assignment via API.
05. Jul 2019 13:52:07 CEST jreher Updated APD location via API.
03. Jan 2019 00:14:28 CET Tobias Corrected dark current unit.
06. Jul 2018 22:38:51 CEST Tobias Imported from Hamamatsu datasheet.