RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 3008033739 - Archived data from Tue, 11. August 2020 18:07:11 CEST


This is all the information about APD 3008033739. If it is wrong, edit the data.

Subdetector specification:  
Serial: 3008033739
Type: Hamamatsu Avalanche Photo Diode
Detector: unassigned
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: none
   
Manufacturer information:  
Wafer position: D10
Break-through voltage: 412 V
Voltage for Gain 100 (T=+25°C): 383.1 V
Dark current: 25.6 nA
   
Screening Logistics:  
Available: Yes
Storage Box: 118
Position in Box: 24
EP1 batch: 126
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: none
Sent for analysis after irradiation: none
Return for assembly: none
   
Irradiation:  
Date: none
Dose used: none
Temperature: none
Position: none
Bias voltage: none     
   
Annealing:  
Date: none
Temperature: none
Duration: none
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 383.0919871 V     T = -25 °C: 347.5886433 V
Voltage for Gain 150: T = +20 °C: 390.9049783 V     T = -25 °C: 354.9332637 V
Voltage for Gain 200: T = +20 °C: 395.2533905 V     T = -25 °C: 359.0446987 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.597719992 V-1     T = -25 °C: 4.72487214 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 9.111690731 V-1     T = -25 °C: 9.629012633 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 14.21328487 V-1     T = -25 °C: 15.28065516 V-1
Break-through voltage: T = +20 °C: 410.8485759 V     T = -25 °C: 374.2752441 V
   
Notes:


Version history

Time Author Change comment
05. Feb 2021 11:03:37 CET jreher Updated box and position via API.
11. Aug 2020 18:07:11 CEST jreher Updated APD characteristics via API.
11. Aug 2020 17:49:41 CEST jreher Updated APD location via API.
02. Mar 2020 16:21:36 CET vwerner Updated APD batch assignment via API.
29. Nov 2019 11:47:17 CET jreher Updated APD location via API.
27. Nov 2019 19:49:33 CET jreher Updated APD location via API.
25. Nov 2019 17:20:18 CET jreher Updated box and position via API.
30. Aug 2019 11:28:11 CEST jgrochowski Updated APD batch assignment via API.
05. Jul 2019 13:52:05 CEST jreher Updated APD location via API.
03. Jan 2019 00:14:27 CET Tobias Corrected dark current unit.
06. Jul 2018 22:38:51 CEST Tobias Imported from Hamamatsu datasheet.