RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 3002033315 - Archived data from Wed, 16. June 2021 16:00:50 CEST


This is all the information about APD 3002033315. If it is wrong, edit the data.

Subdetector specification:  
Serial: 3002033315
Type: Hamamatsu Avalanche Photo Diode
Detector: unassigned
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: none
   
Manufacturer information:  
Wafer position: D07
Break-through voltage: 415 V
Voltage for Gain 100 (T=+25°C): 386.3 V
Dark current: 20 nA
   
Screening Logistics:  
Available: Yes
Storage Box: 112
Position in Box: 28
EP1 batch: 117
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: none
Sent for analysis after irradiation: none
Return for assembly: none
   
Irradiation:  
Date: none
Dose used: none
Temperature: none
Position: none
Bias voltage: none     
   
Annealing:  
Date: none
Temperature: none
Duration: none
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 386.9923129 V     T = -25 °C: 351.0749864 V
Voltage for Gain 150: T = +20 °C: 394.824194 V     T = -25 °C: 358.4913423 V
Voltage for Gain 200: T = +20 °C: 399.177136 V     T = -25 °C: 362.6261124 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.58182327 V-1     T = -25 °C: 4.845475249 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 9.067044148 V-1     T = -25 °C: 9.890060627 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 14.12047236 V-1     T = -25 °C: 15.78814337 V-1
Break-through voltage: T = +20 °C: 414.5491047 V     T = -25 °C: 377.7855571 V
   
Notes:


Version history

Time Author Change comment
19. Nov 2021 13:05:35 CET oafedulidis Assigned to capsule by script
16. Jun 2021 16:04:40 CEST oafedulidis Updated box and position via API.
16. Jun 2021 16:00:50 CEST oafedulidis Updated box and position via API.
16. Jun 2021 15:53:35 CEST oafedulidis Updated box and position via API.
05. Feb 2021 11:02:50 CET jreher Updated box and position via API.
11. Aug 2020 17:46:46 CEST jreher Updated APD characteristics via API.
03. Dec 2019 15:03:11 CET jreher Updated APD location via API.
29. Nov 2019 11:46:54 CET jreher Updated APD location via API.
27. Nov 2019 19:48:26 CET jreher Updated APD location via API.
25. Nov 2019 17:19:47 CET jreher Updated box and position via API.
14. Nov 2019 17:31:22 CET spankonin Updated APD batch assignment via API.
27. Aug 2019 12:11:46 CEST jgrochowski Updated APD batch assignment via API.
05. Jul 2019 13:51:18 CEST jreher Updated APD location via API.
03. Jan 2019 00:14:18 CET Tobias Corrected dark current unit.
06. Jul 2018 22:38:45 CEST Tobias Imported from Hamamatsu datasheet.