Serial: 2903032657
Type: Hamamatsu Avalanche Photo Diode
Detector: unassigned
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: none
   
Manufacturer information:  
Wafer position: D04
Break-through voltage: 407 V
Voltage for Gain 100 (T=+25°C): 378.1 V
Dark current: 6.1 nA
   
Screening Logistics:  
Available: Yes
Storage Box: 108
Position in Box: 50
EP1 batch: 111
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: none
Sent for analysis after irradiation: none
Return for assembly: none
   
Irradiation:  
Date: none
Dose used: none
Temperature: none
Position: none
Bias voltage: none     
   
Annealing:  
Date: none
Temperature: none
Duration: none
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 378.1931746 V     T = -25 °C: 343.2972445 V
Voltage for Gain 150: T = +20 °C: 386.0373677 V     T = -25 °C: 350.6484996 V
Voltage for Gain 200: T = +20 °C: 390.4293164 V     T = -25 °C: 354.768268 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.530803488 V-1     T = -25 °C: 4.802798548 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 8.797750857 V-1     T = -25 °C: 9.765706151 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 15.21469383 V-1     T = -25 °C: 15.43470711 V-1
Break-through voltage: T = +20 °C: 406.2642445 V     T = -25 °C: 370.147721 V
   
Notes:


Version history

Time Author Change comment
29. Mar 2022 12:39:09 CEST oafedulidis Assigned to capsule by script
16. Jun 2021 16:03:09 CEST oafedulidis Updated box and position via API.
16. Jun 2021 15:59:34 CEST oafedulidis Updated box and position via API.
16. Jun 2021 15:52:11 CEST oafedulidis Updated box and position via API.
05. Feb 2021 11:02:21 CET jreher Updated box and position via API.
11. Aug 2020 18:06:25 CEST jreher Updated APD characteristics via API.
11. Aug 2020 17:49:07 CEST jreher Updated APD location via API.
25. Feb 2020 17:02:06 CET vwerner Updated APD batch assignment via API.
29. Nov 2019 11:46:41 CET jreher Updated APD location via API.
28. Nov 2019 12:01:17 CET jreher Updated APD location via API.
25. Nov 2019 17:19:34 CET jreher Updated box and position via API.
23. Aug 2019 12:00:04 CEST jgrochowski Updated APD batch assignment via API.
05. Jul 2019 13:50:00 CEST jreher Updated APD location via API.
03. Jan 2019 00:14:12 CET Tobias Corrected dark current unit.
06. Jul 2018 22:38:41 CEST Tobias Imported from Hamamatsu datasheet.