Show Hamamatsu Avalanche Photo Diode 2810031572 - Archived data from Wed, 03. February 2021 18:25:23 CET


This is all the information about APD 2810031572. If it is wrong, edit the data.

Subdetector specification:  
Serial: 2810031572
Type: Hamamatsu Avalanche Photo Diode
Detector: unassigned
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: none
   
Manufacturer information:  
Wafer position: E07
Break-through voltage: 407 V
Voltage for Gain 100 (T=+25°C): 378.1 V
Dark current: 6 nA
   
Screening Logistics:  
Available: Yes
Storage Box: 90
Position in Box: 8
EP1 batch: 80
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: none
Sent for analysis after irradiation: none
Return for assembly: none
   
Irradiation:  
Date: none
Dose used: none
Temperature: none
Position: none
Bias voltage: none     
   
Annealing:  
Date: none
Temperature: none
Duration: none
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 377.8129388 V     T = -25 °C: 343.4746042 V
Voltage for Gain 150: T = +20 °C: 385.64537 V     T = -25 °C: 350.6855695 V
Voltage for Gain 200: T = +20 °C: 390.0145677 V     T = -25 °C: 354.7573222 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.733339593 V-1     T = -25 °C: 4.823788562 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 9.319698124 V-1     T = -25 °C: 10.22508471 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 14.59821905 V-1     T = -25 °C: 16.37476508 V-1
Break-through voltage: T = +20 °C: 405.8663317 V     T = -25 °C: 369.9438767 V
   
Notes:


Version history

Time Author Change comment
05. Feb 2021 10:59:54 CET jreher Updated box and position via API.
03. Feb 2021 18:25:23 CET jreher Updated box and position via API.
11. Aug 2020 18:04:27 CEST jreher Updated APD characteristics via API.
11. Aug 2020 17:54:56 CEST jreher Updated APD location via API.
17. Feb 2020 17:57:00 CET spankonin Updated APD batch assignment via API.
29. Nov 2019 11:50:26 CET jreher Updated APD location via API.
27. Nov 2019 19:44:47 CET jreher Updated APD location via API.
25. Nov 2019 17:18:03 CET jreher Updated box and position via API.
20. Aug 2019 12:17:16 CEST jreher Updated APD batch assignment via API.
05. Jul 2019 13:48:12 CEST jreher Updated APD location via API.
03. Jan 2019 00:13:42 CET Tobias Corrected dark current unit.
20. Jun 2018 13:36:02 CEST Tobias Imported from Hamamatsu datasheet.