RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 2808031440 - Archived data from Tue, 11. August 2020 17:39:28 CEST


This is all the information about APD 2808031440. If it is wrong, edit the data.

Subdetector specification:  
Serial: 2808031440
Type: Hamamatsu Avalanche Photo Diode
Detector: unassigned
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: none
   
Manufacturer information:  
Wafer position: D07
Break-through voltage: 411 V
Voltage for Gain 100 (T=+25°C): 382.2 V
Dark current: 8.6 nA
   
Screening Logistics:  
Available: Yes
Storage Box: 88
Position in Box: 24
EP1 batch: 77
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: 26. Aug 2019
Sent for analysis after irradiation: 02. Sep 2019
Return for assembly: none
   
Irradiation:  
Date: 28. Aug 2019
Dose used: 37 Gy
Temperature: 20 °C
Position: none
Bias voltage: none     
   
Annealing:  
Date: 31. Aug 2019
Temperature: 80 °C
Duration: 48 h
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 382.3747972 V     T = -25 °C: 346.8160754 V
Voltage for Gain 150: T = +20 °C: 390.1940098 V     T = -25 °C: 354.1480728 V
Voltage for Gain 200: T = +20 °C: 394.5459908 V     T = -25 °C: 358.2653353 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.513568525 V-1     T = -25 °C: 5.01618676 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 8.858924822 V-1     T = -25 °C: 9.420987094 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 15.55301202 V-1     T = -25 °C: 14.93118744 V-1
Break-through voltage: T = +20 °C: 410.3538525 V     T = -25 °C: 373.6526183 V
   
Notes:


Version history

Time Author Change comment
11. Jul 2022 14:24:13 CEST oafedulidis Assigned to capsule by script
17. May 2021 09:14:52 CEST oafedulidis Updated box and position via API.
05. Feb 2021 10:59:40 CET jreher Updated box and position via API.
03. Feb 2021 18:25:09 CET jreher Updated box and position via API.
11. Aug 2020 17:39:28 CEST jreher Updated APD characteristics via API.
27. Nov 2019 19:44:26 CET jreher Updated APD location via API.
25. Nov 2019 17:17:55 CET jreher Updated box and position via API.
13. Sep 2019 10:30:23 CEST jgrochowski Updated APD batch assignment via API.
04. Sep 2019 12:48:09 CEST markus.moritz
25. Jul 2019 12:11:10 CEST spankonin Updated APD batch assignment via API.
05. Jul 2019 13:47:57 CEST jreher Updated APD location via API.
03. Jan 2019 00:13:39 CET Tobias Corrected dark current unit.
20. Jun 2018 13:35:57 CEST Tobias Imported from Hamamatsu datasheet.