RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 2805031187 - Archived data from Tue, 11. August 2020 17:39:14 CEST


This is all the information about APD 2805031187. If it is wrong, edit the data.

Subdetector specification:  
Serial: 2805031187
Type: Hamamatsu Avalanche Photo Diode
Detector: assigned to 2408026530/2805031187
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: red
   
Manufacturer information:  
Wafer position: B06
Break-through voltage: 410 V
Voltage for Gain 100 (T=+25°C): 384.5 V
Dark current: 5.9 nA
   
Screening Logistics:  
Available: No
Storage Box: 86
Position in Box: 32
EP1 batch: 75
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: 26. Aug 2019
Sent for analysis after irradiation: 02. Sep 2019
Return for assembly: none
   
Irradiation:  
Date: 28. Aug 2019
Dose used: 37 Gy
Temperature: 20 °C
Position: none
Bias voltage: none     
   
Annealing:  
Date: 31. Aug 2019
Temperature: 80 °C
Duration: 48 h
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 384.5702452 V     T = -25 °C: 349.0780991 V
Voltage for Gain 150: T = +20 °C: 392.3902761 V     T = -25 °C: 356.5182137 V
Voltage for Gain 200: T = +20 °C: 396.7310429 V     T = -25 °C: 360.7002722 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.469191753 V-1     T = -25 °C: 4.889210449 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 8.775171672 V-1     T = -25 °C: 9.045969236 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 15.51112956 V-1     T = -25 °C: 16.16218362 V-1
Break-through voltage: T = +20 °C: 411.5112532 V     T = -25 °C: 375.2819564 V
   
Notes:


Version history

Time Author Change comment
05. Feb 2021 10:59:27 CET jreher Updated box and position via API.
03. Feb 2021 18:24:56 CET jreher Updated box and position via API.
11. Aug 2020 17:39:14 CEST jreher Updated APD characteristics via API.
25. Jun 2020 11:14:25 CEST jreher Assigned to capsule by script
20. Feb 2020 13:34:41 CET jreher Updated box and position via API.
27. Nov 2019 19:44:06 CET jreher Updated APD location via API.
25. Nov 2019 17:17:44 CET jreher Updated box and position via API.
13. Sep 2019 10:29:40 CEST jgrochowski Updated APD batch assignment via API.
04. Sep 2019 13:41:38 CEST markus.moritz
25. Jul 2019 12:11:42 CEST spankonin Updated APD batch assignment via API.
05. Jul 2019 13:47:43 CEST jreher Updated APD location via API.
03. Jan 2019 00:13:36 CET Tobias Corrected dark current unit.
24. May 2018 14:39:23 CEST Tobias Updated location to "Bochum".
18. May 2018 12:05:52 CEST Tobias Imported from Hamamatsu datasheet.