RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 2802030920 - Archived data from Thu, 03. September 2020 14:25:34 CEST


This is all the information about APD 2802030920. If it is wrong, edit the data.

Subdetector specification:  
Serial: 2802030920
Type: Hamamatsu Avalanche Photo Diode
Detector: assigned to 2802030920/0821008841
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: blue
   
Manufacturer information:  
Wafer position: C05
Break-through voltage: 409 V
Voltage for Gain 100 (T=+25°C): 379.7 V
Dark current: 8.4 nA
   
Screening Logistics:  
Available: No
Storage Box: 84
Position in Box: 17
EP1 batch: 71
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: 26. Aug 2019
Sent for analysis after irradiation: 02. Sep 2019
Return for assembly: none
   
Irradiation:  
Date: 28. Aug 2019
Dose used: 37 Gy
Temperature: 20 °C
Position: none
Bias voltage: none     
   
Annealing:  
Date: 31. Aug 2019
Temperature: 80 °C
Duration: 48 h
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 379.7525039 V     T = -25 °C: 344.6998878 V
Voltage for Gain 150: T = +20 °C: 387.5843737 V     T = -25 °C: 352.0383732 V
Voltage for Gain 200: T = +20 °C: 391.9480769 V     T = -25 °C: 356.1694629 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.704639046 V-1     T = -25 °C: 5.039594115 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 9.281041797 V-1     T = -25 °C: 9.425239486 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 14.39636242 V-1     T = -25 °C: 14.78701575 V-1
Break-through voltage: T = +20 °C: 407.7573132 V     T = -25 °C: 371.2958998 V
   
Notes:


Version history

Time Author Change comment
05. Feb 2021 10:59:10 CET jreher Updated box and position via API.
03. Feb 2021 18:24:39 CET jreher Updated box and position via API.
03. Sep 2020 14:25:34 CEST jreher Assigned to capsule by script
11. Aug 2020 17:38:53 CEST jreher Updated APD characteristics via API.
13. Feb 2020 14:23:31 CET jreher Updated box and position via API.
27. Nov 2019 19:43:43 CET jreher Updated APD location via API.
25. Nov 2019 17:17:33 CET jreher Updated box and position via API.
13. Sep 2019 10:28:00 CEST jgrochowski Updated APD batch assignment via API.
04. Sep 2019 12:59:50 CEST markus.moritz
25. Jul 2019 12:12:45 CEST spankonin Updated APD batch assignment via API.
05. Jul 2019 13:47:26 CEST jreher Updated APD location via API.
03. Jan 2019 00:13:31 CET Tobias Corrected dark current unit.
24. May 2018 14:39:23 CEST Tobias Updated location to "Bochum".
18. May 2018 12:05:06 CEST Tobias Imported from Hamamatsu datasheet.