RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 2710030755 - Archived data from Mon, 03. May 2021 10:27:10 CEST


This is all the information about APD 2710030755. If it is wrong, edit the data.

Subdetector specification:  
Serial: 2710030755
Type: Hamamatsu Avalanche Photo Diode
Detector: unassigned
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: none
   
Manufacturer information:  
Wafer position: C09
Break-through voltage: 412 V
Voltage for Gain 100 (T=+25°C): 383.9 V
Dark current: 14.4 nA
   
Screening Logistics:  
Available: Yes
Storage Box: 82
Position in Box: 15
EP1 batch: 68
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: 26. Aug 2019
Sent for analysis after irradiation: 02. Sep 2019
Return for assembly: none
   
Irradiation:  
Date: 28. Aug 2019
Dose used: 37 Gy
Temperature: 20 °C
Position: none
Bias voltage: none     
   
Annealing:  
Date: 31. Aug 2019
Temperature: 80 °C
Duration: 48 h
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 384.0539876 V     T = -25 °C: 348.3989894 V
Voltage for Gain 150: T = +20 °C: 391.9656647 V     T = -25 °C: 355.842282 V
Voltage for Gain 200: T = +20 °C: 396.3789653 V     T = -25 °C: 360.0024921 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.525236123 V-1     T = -25 °C: 4.698499507 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 8.78847276 V-1     T = -25 °C: 9.428345101 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 15.28637725 V-1     T = -25 °C: 15.00188548 V-1
Break-through voltage: T = +20 °C: 411.4050987 V     T = -25 °C: 374.8338711 V
   
Notes:


Version history

Time Author Change comment
27. Jun 2022 12:43:53 CEST oafedulidis Assigned to capsule by script
27. Aug 2021 11:14:06 CEST oafedulidis Updated box and position via API.
03. May 2021 10:27:10 CEST oafedulidis Updated box and position via API.
05. Feb 2021 10:58:54 CET jreher Updated box and position via API.
03. Feb 2021 18:24:25 CET jreher Updated box and position via API.
11. Aug 2020 17:38:34 CEST jreher Updated APD characteristics via API.
27. Nov 2019 19:43:21 CET jreher Updated APD location via API.
25. Nov 2019 17:17:24 CET jreher Updated box and position via API.
10. Sep 2019 11:22:20 CEST jgrochowski Updated APD batch assignment via API.
04. Sep 2019 11:45:16 CEST markus.moritz
25. Jul 2019 12:13:32 CEST spankonin Updated APD batch assignment via API.
05. Jul 2019 13:47:10 CEST jreher Updated APD location via API.
03. Jan 2019 00:13:28 CET Tobias Corrected dark current unit.
24. May 2018 14:39:23 CEST Tobias Updated location to "Bochum".
18. May 2018 12:04:29 CEST Tobias Imported from Hamamatsu datasheet.