RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 2707030538 - Archived data from Fri, 05. February 2021 10:58:30 CET


This is all the information about APD 2707030538. If it is wrong, edit the data.

Subdetector specification:  
Serial: 2707030538
Type: Hamamatsu Avalanche Photo Diode
Detector: unassigned
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: none
   
Manufacturer information:  
Wafer position: F11
Break-through voltage: 411 V
Voltage for Gain 100 (T=+25°C): 383.3 V
Dark current: 8.2 nA
   
Screening Logistics:  
Available: Yes
Storage Box: 79
Position in Box: 14
EP1 batch: 63
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: 26. Aug 2019
Sent for analysis after irradiation: 02. Sep 2019
Return for assembly: none
   
Irradiation:  
Date: 28. Aug 2019
Dose used: 37 Gy
Temperature: 20 °C
Position: none
Bias voltage: none     
   
Annealing:  
Date: 31. Aug 2019
Temperature: 80 °C
Duration: 48 h
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 383.4215236 V     T = -25 °C: 347.8011978 V
Voltage for Gain 150: T = +20 °C: 391.2204456 V     T = -25 °C: 355.1899634 V
Voltage for Gain 200: T = +20 °C: 395.5540232 V     T = -25 °C: 359.3401217 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.472087148 V-1     T = -25 °C: 5.028370612 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 8.850661008 V-1     T = -25 °C: 9.298483306 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 15.56906155 V-1     T = -25 °C: 14.6496594 V-1
Break-through voltage: T = +20 °C: 410.2864382 V     T = -25 °C: 374.056425 V
   
Notes:


Version history

Time Author Change comment
16. Feb 2021 12:25:43 CET jreher Updated box and position via API.
15. Feb 2021 15:40:05 CET jreher Assigned to capsule by script
05. Feb 2021 10:58:30 CET jreher Updated box and position via API.
03. Feb 2021 18:24:00 CET jreher Updated box and position via API.
11. Aug 2020 17:38:05 CEST jreher Updated APD characteristics via API.
27. Nov 2019 19:42:47 CET jreher Updated APD location via API.
25. Nov 2019 17:17:10 CET jreher Updated box and position via API.
10. Sep 2019 11:19:58 CEST jgrochowski Updated APD batch assignment via API.
04. Sep 2019 10:22:11 CEST markus.moritz
25. Jul 2019 12:14:40 CEST spankonin Updated APD batch assignment via API.
05. Jul 2019 13:46:46 CEST jreher Updated APD location via API.
03. Jan 2019 00:13:23 CET Tobias Corrected dark current unit.
24. May 2018 14:39:23 CEST Tobias Updated location to "Bochum".
18. May 2018 12:03:28 CEST Tobias Imported from Hamamatsu datasheet.