RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 2707030518 - Archived data from Mon, 03. May 2021 10:26:08 CEST


This is all the information about APD 2707030518. If it is wrong, edit the data.

Subdetector specification:  
Serial: 2707030518
Type: Hamamatsu Avalanche Photo Diode
Detector: unassigned
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: none
   
Manufacturer information:  
Wafer position: A08
Break-through voltage: 412 V
Voltage for Gain 100 (T=+25°C): 383.4 V
Dark current: 6.1 nA
   
Screening Logistics:  
Available: Yes
Storage Box: 93
Position in Box: 33
EP1 batch: 63
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: 26. Aug 2019
Sent for analysis after irradiation: 02. Sep 2019
Return for assembly: none
   
Irradiation:  
Date: 28. Aug 2019
Dose used: 37 Gy
Temperature: 20 °C
Position: none
Bias voltage: none     
   
Annealing:  
Date: 31. Aug 2019
Temperature: 80 °C
Duration: 48 h
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 383.480713 V     T = -25 °C: 348.0664619 V
Voltage for Gain 150: T = +20 °C: 391.3777008 V     T = -25 °C: 355.5421546 V
Voltage for Gain 200: T = +20 °C: 395.7692521 V     T = -25 °C: 359.7268815 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.409579629 V-1     T = -25 °C: 4.854270288 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 8.54960227 V-1     T = -25 °C: 9.804963946 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 14.75896809 V-1     T = -25 °C: 15.43551566 V-1
Break-through voltage: T = +20 °C: 411.2649564 V     T = -25 °C: 375.1527679 V
   
Notes:


Version history

Time Author Change comment
15. Nov 2022 13:09:06 CET oafedulidis Assigned to capsule by script
30. Aug 2021 14:07:11 CEST oafedulidis Updated box and position via API.
03. May 2021 10:26:08 CEST oafedulidis Updated box and position via API.
05. Feb 2021 10:58:29 CET jreher Updated box and position via API.
03. Feb 2021 18:23:59 CET jreher Updated box and position via API.
11. Aug 2020 17:38:04 CEST jreher Updated APD characteristics via API.
27. Nov 2019 19:42:47 CET jreher Updated APD location via API.
25. Nov 2019 17:17:09 CET jreher Updated box and position via API.
10. Sep 2019 11:19:57 CEST jgrochowski Updated APD batch assignment via API.
04. Sep 2019 10:19:19 CEST markus.moritz
25. Jul 2019 12:14:40 CEST spankonin Updated APD batch assignment via API.
05. Jul 2019 13:46:45 CEST jreher Updated APD location via API.
03. Jan 2019 00:13:22 CET Tobias Corrected dark current unit.
24. May 2018 14:39:23 CEST Tobias Updated location to "Bochum".
18. May 2018 12:03:25 CEST Tobias Imported from Hamamatsu datasheet.