RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 2706030453 - Archived data from Tue, 11. August 2020 17:38:01 CEST


This is all the information about APD 2706030453. If it is wrong, edit the data.

Subdetector specification:  
Serial: 2706030453
Type: Hamamatsu Avalanche Photo Diode
Detector: unassigned
Unit: unassigned
Preamp: 0
Current location: Giessen
   
Installation information:  
Label: none
   
Manufacturer information:  
Wafer position: C06
Break-through voltage: 416 V
Voltage for Gain 100 (T=+25°C): 387.8 V
Dark current: 8.3 nA
   
Screening Logistics:  
Available: Yes
Storage Box: 78
Position in Box: 27
EP1 batch: 62
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: 26. Aug 2019
Sent for analysis after irradiation: 02. Sep 2019
Return for assembly: none
   
Irradiation:  
Date: 28. Aug 2019
Dose used: 37 Gy
Temperature: 20 °C
Position: none
Bias voltage: none     
   
Annealing:  
Date: 31. Aug 2019
Temperature: 80 °C
Duration: 48 h
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 388.0421438 V     T = -25 °C: 352.2140174 V
Voltage for Gain 150: T = +20 °C: 395.8588866 V     T = -25 °C: 359.6742231 V
Voltage for Gain 200: T = +20 °C: 400.2022688 V     T = -25 °C: 363.8413437 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.622338617 V-1     T = -25 °C: 4.806392119 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 9.174787568 V-1     T = -25 °C: 9.773414439 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 14.35323565 V-1     T = -25 °C: 15.50659937 V-1
Break-through voltage: T = +20 °C: 415.1401738 V     T = -25 °C: 378.8467863 V
   
Notes:


Version history

Time Author Change comment
16. Nov 2021 09:35:19 CET oafedulidis Assigned to capsule by script
10. May 2021 14:01:00 CEST oafedulidis Updated box and position via API.
05. Feb 2021 10:58:23 CET jreher Updated box and position via API.
03. Feb 2021 18:23:54 CET jreher Updated box and position via API.
11. Aug 2020 17:54:53 CEST jreher Updated APD location via API.
11. Aug 2020 17:38:01 CEST jreher Updated APD characteristics via API.
29. Nov 2019 11:50:23 CET jreher Updated APD location via API.
27. Nov 2019 19:42:37 CET jreher Updated APD location via API.
25. Nov 2019 17:17:06 CET jreher Updated box and position via API.
06. Sep 2019 13:13:36 CEST spankonin Updated APD batch assignment via API.
03. Sep 2019 15:01:50 CEST markus.moritz
08. Jul 2019 14:56:46 CEST spankonin Updated APD batch assignment via API.
05. Jul 2019 13:46:39 CEST jreher Updated APD location via API.
03. Jan 2019 00:13:21 CET Tobias Corrected dark current unit.
24. May 2018 14:39:23 CEST Tobias Updated location to "Bochum".
18. May 2018 12:03:14 CEST Tobias Imported from Hamamatsu datasheet.