RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 2706030449 - Archived data from Mon, 16. May 2022 13:33:54 CEST


This is all the information about APD 2706030449. If it is wrong, edit the data.

Subdetector specification:  
Serial: 2706030449
Type: Hamamatsu Avalanche Photo Diode
Detector: unassigned
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: none
   
Manufacturer information:  
Wafer position: E11
Break-through voltage: 407 V
Voltage for Gain 100 (T=+25°C): 378.4 V
Dark current: 12.2 nA
   
Screening Logistics:  
Available: Yes
Storage Box: 78
Position in Box: 23
EP1 batch: 62
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: 26. Aug 2019
Sent for analysis after irradiation: 02. Sep 2019
Return for assembly: none
   
Irradiation:  
Date: 28. Aug 2019
Dose used: 37 Gy
Temperature: 20 °C
Position: none
Bias voltage: none     
   
Annealing:  
Date: 31. Aug 2019
Temperature: 80 °C
Duration: 48 h
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 378.4183614 V     T = -25 °C: 343.2863977 V
Voltage for Gain 150: T = +20 °C: 386.2273641 V     T = -25 °C: 350.6277602 V
Voltage for Gain 200: T = +20 °C: 390.5867402 V     T = -25 °C: 354.7254866 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.507248934 V-1     T = -25 °C: 4.842406471 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 8.81879951 V-1     T = -25 °C: 9.994066464 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 15.38066969 V-1     T = -25 °C: 15.99161099 V-1
Break-through voltage: T = +20 °C: 406.20524 V     T = -25 °C: 369.9526525 V
   
Notes:


Version history

Time Author Change comment
16. Nov 2022 11:27:06 CET oafedulidis Assigned to capsule by script
16. May 2022 13:33:54 CEST oafedulidis Updated box and position via API.
28. Apr 2021 13:29:13 CEST oafedulidis Updated box and position via API.
05. Feb 2021 10:58:23 CET jreher Updated box and position via API.
03. Feb 2021 18:23:53 CET jreher Updated box and position via API.
11. Aug 2020 17:54:53 CEST jreher Updated APD location via API.
11. Aug 2020 17:38:00 CEST jreher Updated APD characteristics via API.
29. Nov 2019 11:50:23 CET jreher Updated APD location via API.
27. Nov 2019 19:42:37 CET jreher Updated APD location via API.
25. Nov 2019 17:17:06 CET jreher Updated box and position via API.
06. Sep 2019 13:13:36 CEST spankonin Updated APD batch assignment via API.
03. Sep 2019 15:00:28 CEST markus.moritz
08. Jul 2019 14:56:46 CEST spankonin Updated APD batch assignment via API.
05. Jul 2019 13:46:38 CEST jreher Updated APD location via API.
03. Jan 2019 00:13:21 CET Tobias Corrected dark current unit.
24. May 2018 14:39:23 CEST Tobias Updated location to "Bochum".
18. May 2018 12:03:13 CEST Tobias Imported from Hamamatsu datasheet.