RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 2706030430 - Archived data from Fri, 05. February 2021 10:58:22 CET


This is all the information about APD 2706030430. If it is wrong, edit the data.

Subdetector specification:  
Serial: 2706030430
Type: Hamamatsu Avalanche Photo Diode
Detector: unassigned
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: none
   
Manufacturer information:  
Wafer position: A09
Break-through voltage: 415 V
Voltage for Gain 100 (T=+25°C): 387.2 V
Dark current: 6.6 nA
   
Screening Logistics:  
Available: Yes
Storage Box: 78
Position in Box: 13
EP1 batch: 62
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: 26. Aug 2019
Sent for analysis after irradiation: 02. Sep 2019
Return for assembly: none
   
Irradiation:  
Date: 28. Aug 2019
Dose used: 37 Gy
Temperature: 20 °C
Position: none
Bias voltage: none     
   
Annealing:  
Date: 31. Aug 2019
Temperature: 80 °C
Duration: 48 h
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 386.821323 V     T = -25 °C: 351.0140536 V
Voltage for Gain 150: T = +20 °C: 394.7010617 V     T = -25 °C: 358.5444445 V
Voltage for Gain 200: T = +20 °C: 399.0798075 V     T = -25 °C: 362.7522645 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.626474523 V-1     T = -25 °C: 4.804797551 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 9.200181908 V-1     T = -25 °C: 9.735858296 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 14.25051798 V-1     T = -25 °C: 15.3523973 V-1
Break-through voltage: T = +20 °C: 414.6544997 V     T = -25 °C: 378.3837033 V
   
Notes:


Version history

Time Author Change comment
25. Oct 2021 14:40:21 CEST oafedulidis Assigned to capsule by script
15. Apr 2021 17:01:12 CEST jreher Updated box and position via API.
05. Feb 2021 10:58:22 CET jreher Updated box and position via API.
03. Feb 2021 18:23:52 CET jreher Updated box and position via API.
11. Aug 2020 17:54:52 CEST jreher Updated APD location via API.
11. Aug 2020 17:37:58 CEST jreher Updated APD characteristics via API.
29. Nov 2019 11:50:22 CET jreher Updated APD location via API.
27. Nov 2019 19:42:36 CET jreher Updated APD location via API.
25. Nov 2019 17:17:05 CET jreher Updated box and position via API.
06. Sep 2019 13:13:35 CEST spankonin Updated APD batch assignment via API.
03. Sep 2019 14:55:45 CEST markus.moritz
08. Jul 2019 14:56:45 CEST spankonin Updated APD batch assignment via API.
05. Jul 2019 13:46:37 CEST jreher Updated APD location via API.
03. Jan 2019 00:13:21 CET Tobias Corrected dark current unit.
24. May 2018 14:39:23 CEST Tobias Updated location to "Bochum".
18. May 2018 12:03:09 CEST Tobias Imported from Hamamatsu datasheet.