RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 2704030266 - Archived data from Fri, 05. February 2021 10:58:04 CET


This is all the information about APD 2704030266. If it is wrong, edit the data.

Subdetector specification:  
Serial: 2704030266
Type: Hamamatsu Avalanche Photo Diode
Detector: unassigned
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: none
   
Manufacturer information:  
Wafer position: D12
Break-through voltage: 410 V
Voltage for Gain 100 (T=+25°C): 381.1 V
Dark current: 17.7 nA
   
Screening Logistics:  
Available: Yes
Storage Box: 76
Position in Box: 16
EP1 batch: 59
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: 26. Aug 2019
Sent for analysis after irradiation: 02. Sep 2019
Return for assembly: none
   
Irradiation:  
Date: 28. Aug 2019
Dose used: 37 Gy
Temperature: 20 °C
Position: none
Bias voltage: none     
   
Annealing:  
Date: 31. Aug 2019
Temperature: 80 °C
Duration: 48 h
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 381.471916 V     T = -25 °C: 346.1948892 V
Voltage for Gain 150: T = +20 °C: 389.2839913 V     T = -25 °C: 353.5307564 V
Voltage for Gain 200: T = +20 °C: 393.6361845 V     T = -25 °C: 357.6394705 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.471400682 V-1     T = -25 °C: 4.888406606 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 8.819633373 V-1     T = -25 °C: 9.092961984 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 15.32395833 V-1     T = -25 °C: 16.23225436 V-1
Break-through voltage: T = +20 °C: 408.9843915 V     T = -25 °C: 372.5668261 V
   
Notes:


Version history

Time Author Change comment
27. Jun 2022 12:42:48 CEST oafedulidis Assigned to capsule by script
16. Jun 2021 16:03:55 CEST oafedulidis Updated box and position via API.
16. Jun 2021 16:00:10 CEST oafedulidis Updated box and position via API.
16. Jun 2021 15:52:51 CEST oafedulidis Updated box and position via API.
05. Feb 2021 10:58:04 CET jreher Updated box and position via API.
03. Feb 2021 18:23:37 CET jreher Updated box and position via API.
11. Aug 2020 17:37:40 CEST jreher Updated APD characteristics via API.
27. Nov 2019 19:42:14 CET jreher Updated APD location via API.
25. Nov 2019 17:16:56 CET jreher Updated box and position via API.
06. Sep 2019 13:12:08 CEST spankonin Updated APD batch assignment via API.
03. Sep 2019 12:17:53 CEST markus.moritz
08. Jul 2019 14:55:55 CEST spankonin Updated APD batch assignment via API.
05. Jul 2019 13:46:22 CEST jreher Updated APD location via API.
03. Jan 2019 00:13:17 CET Tobias Corrected dark current unit.
24. May 2018 14:39:23 CEST Tobias Updated location to "Bochum".
18. May 2018 12:02:29 CEST Tobias Imported from Hamamatsu datasheet.