Show Hamamatsu Avalanche Photo Diode 2703030213 - Archived data from Wed, 03. February 2021 18:23:32 CET
This is all the information about APD 2703030213. If it is wrong, edit the data.
Subdetector specification: | |
---|---|
Serial: | 2703030213 |
Type: | Hamamatsu Avalanche Photo Diode |
Detector: | unassigned |
Unit: | unassigned |
Preamp: | 0 |
Current location: | Bochum |
Installation information: | |
Label: | none |
Manufacturer information: | |
Wafer position: | C08 |
Break-through voltage: | 407 V |
Voltage for Gain 100 (T=+25°C): | 378.7 V |
Dark current: | 11.4 nA |
Screening Logistics: | |
Available: | Yes |
Storage Box: | 75 |
Position in Box: | 38 |
EP1 batch: | 58 |
EP1 batch after irradiation: | none |
Shipment: | |
Grid number: | none |
Position in grid: | none |
Arrival for irradiation: | 26. Aug 2019 |
Sent for analysis after irradiation: | 02. Sep 2019 |
Return for assembly: | none |
Irradiation: | |
Date: | 28. Aug 2019 |
Dose used: | 37 Gy |
Temperature: | 20 °C |
Position: | none |
Bias voltage: | none |
Annealing: | |
Date: | 31. Aug 2019 |
Temperature: | 80 °C |
Duration: | 48 h |
Measurement results: | |
Voltage for Gain 100: | T = +20 °C: 378.8261351 V T = -25 °C: 343.6127872 V |
Voltage for Gain 150: | T = +20 °C: 386.6660424 V T = -25 °C: 350.9990378 V |
Voltage for Gain 200: | T = +20 °C: 391.0323322 V T = -25 °C: 355.119818 V |
Gain/Voltage slope at M = 100: | T = +20 °C: 4.644172642 V-1 T = -25 °C: 4.784427323 V-1 |
Gain/Voltage slope at M = 150: | T = +20 °C: 9.145024368 V-1 T = -25 °C: 9.351396754 V-1 |
Gain/Voltage slope at M = 200: | T = +20 °C: 14.18800946 V-1 T = -25 °C: 14.97802338 V-1 |
Break-through voltage: | T = +20 °C: 406.7313767 V T = -25 °C: 370.4160763 V |
Notes: |
Version history
Time | Author | Change comment |
---|---|---|
20. Oct 2021 09:32:17 CEST | oafedulidis | Assigned to capsule by script |
15. Apr 2021 17:00:51 CEST | jreher | Updated box and position via API. |
05. Feb 2021 10:57:59 CET | jreher | Updated box and position via API. |
03. Feb 2021 18:23:32 CET | jreher | Updated box and position via API. |
11. Aug 2020 17:37:35 CEST | jreher | Updated APD characteristics via API. |
27. Nov 2019 19:42:04 CET | jreher | Updated APD location via API. |
25. Nov 2019 17:16:53 CET | jreher | Updated box and position via API. |
06. Sep 2019 13:11:42 CEST | spankonin | Updated APD batch assignment via API. |
03. Sep 2019 11:49:42 CEST | markus.moritz | |
08. Jul 2019 14:55:40 CEST | spankonin | Updated APD batch assignment via API. |
05. Jul 2019 13:46:15 CEST | jreher | Updated APD location via API. |
03. Jan 2019 00:13:16 CET | Tobias | Corrected dark current unit. |
24. May 2018 14:39:23 CEST | Tobias | Updated location to "Bochum". |
18. May 2018 12:02:18 CEST | Tobias | Imported from Hamamatsu datasheet. |