RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 2703030186 - Archived data from Mon, 30. August 2021 14:07:11 CEST


This is all the information about APD 2703030186. If it is wrong, edit the data.

Subdetector specification:  
Serial: 2703030186
Type: Hamamatsu Avalanche Photo Diode
Detector: unassigned
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: none
   
Manufacturer information:  
Wafer position: D12
Break-through voltage: 412 V
Voltage for Gain 100 (T=+25°C): 384.1 V
Dark current: 14.4 nA
   
Screening Logistics:  
Available: Yes
Storage Box: 178
Position in Box: 36
EP1 batch: 57
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: 26. Aug 2019
Sent for analysis after irradiation: 02. Sep 2019
Return for assembly: none
   
Irradiation:  
Date: 28. Aug 2019
Dose used: 37 Gy
Temperature: 20 °C
Position: none
Bias voltage: none     
   
Annealing:  
Date: 31. Aug 2019
Temperature: 80 °C
Duration: 48 h
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 383.8802097 V     T = -25 °C: 348.1430327 V
Voltage for Gain 150: T = +20 °C: 391.7054692 V     T = -25 °C: 355.5855699 V
Voltage for Gain 200: T = +20 °C: 396.0417391 V     T = -25 °C: 359.7497091 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.70201412 V-1     T = -25 °C: 4.842121431 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 9.336281294 V-1     T = -25 °C: 8.967161973 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 14.66144925 V-1     T = -25 °C: 15.80186439 V-1
Break-through voltage: T = +20 °C: 411.161567 V     T = -25 °C: 374.8272023 V
   
Notes:


Version history

Time Author Change comment
15. Nov 2022 13:09:08 CET oafedulidis Assigned to capsule by script
30. Aug 2021 14:07:11 CEST oafedulidis Updated box and position via API.
03. May 2021 10:24:49 CEST oafedulidis Updated box and position via API.
05. Feb 2021 10:57:56 CET jreher Updated box and position via API.
03. Feb 2021 18:23:29 CET jreher Updated box and position via API.
11. Aug 2020 17:37:31 CEST jreher Updated APD characteristics via API.
27. Nov 2019 19:42:03 CET jreher Updated APD location via API.
25. Nov 2019 17:16:51 CET jreher Updated box and position via API.
06. Sep 2019 13:11:22 CEST spankonin Updated APD batch assignment via API.
03. Sep 2019 11:35:53 CEST markus.moritz
08. Jul 2019 14:55:24 CEST spankonin Updated APD batch assignment via API.
05. Jul 2019 13:46:14 CEST jreher Updated APD location via API.
03. Jan 2019 00:13:16 CET Tobias Corrected dark current unit.
24. May 2018 14:39:23 CEST Tobias Updated location to "Bochum".
18. May 2018 12:02:12 CEST Tobias Imported from Hamamatsu datasheet.