RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 2701030023 - Archived data from Thu, 03. January 2019 00:13:12 CET


This is all the information about APD 2701030023. If it is wrong, edit the data.

Subdetector specification:  
Serial: 2701030023
Type: Hamamatsu Avalanche Photo Diode
Detector: unassigned
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: none
   
Manufacturer information:  
Wafer position: E11
Break-through voltage: 412 V
Voltage for Gain 100 (T=+25°C): 383.7 V
Dark current: 10.1 nA
   
Screening Logistics:  
Available: Yes
Storage Box: 73
Position in Box: 24
EP1 batch: none
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: none
Sent for analysis after irradiation: none
Return for assembly: none
   
Irradiation:  
Date: none
Dose used: none
Temperature: none
Position: none
Bias voltage: none     
   
Annealing:  
Date: none
Temperature: none
Duration: none
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: none     T = -25 °C: none
Voltage for Gain 150: T = +20 °C: none     T = -25 °C: none
Voltage for Gain 200: T = +20 °C: none     T = -25 °C: none
Gain/Voltage slope at M = 100: T = +20 °C: none     T = -25 °C: none
Gain/Voltage slope at M = 150: T = +20 °C: none     T = -25 °C: none
Gain/Voltage slope at M = 200: T = +20 °C: none     T = -25 °C: none
Break-through voltage: T = +20 °C: none     T = -25 °C: none
   
Notes:


Version history

Time Author Change comment
27. Jun 2022 12:43:44 CEST oafedulidis Assigned to capsule by script
27. Aug 2021 11:14:04 CEST oafedulidis Updated box and position via API.
05. Feb 2021 10:57:42 CET jreher Updated box and position via API.
03. Feb 2021 18:23:15 CET jreher Updated box and position via API.
11. Aug 2020 17:37:13 CEST jreher Updated APD characteristics via API.
27. Nov 2019 19:41:41 CET jreher Updated APD location via API.
25. Nov 2019 17:16:42 CET jreher Updated box and position via API.
06. Sep 2019 13:10:03 CEST spankonin Updated APD batch assignment via API.
02. Sep 2019 14:24:01 CEST markus.moritz
08. Jul 2019 14:54:22 CEST spankonin Updated APD batch assignment via API.
05. Jul 2019 13:45:58 CEST jreher Updated APD location via API.
03. Jan 2019 00:13:12 CET Tobias Corrected dark current unit.
24. May 2018 14:39:23 CEST Tobias Updated location to "Bochum".
18. May 2018 12:01:47 CEST Tobias Imported from Hamamatsu datasheet.