RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 2520029116 - Archived data from Fri, 05. February 2021 11:25:13 CET


This is all the information about APD 2520029116. If it is wrong, edit the data.

Subdetector specification:  
Serial: 2520029116
Type: Hamamatsu Avalanche Photo Diode
Detector: unassigned
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: none
   
Manufacturer information:  
Wafer position: E13
Break-through voltage: 407 V
Voltage for Gain 100 (T=+25°C): 377.8 V
Dark current: 6.2 nA
   
Screening Logistics:  
Available: Yes
Storage Box: 286
Position in Box: 30
EP1 batch: 103
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: none
Sent for analysis after irradiation: none
Return for assembly: none
   
Irradiation:  
Date: none
Dose used: none
Temperature: none
Position: none
Bias voltage: none     
   
Annealing:  
Date: none
Temperature: none
Duration: none
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 377.8937047 V     T = -25 °C: 342.7324849 V
Voltage for Gain 150: T = +20 °C: 385.7578786 V     T = -25 °C: 350.1021521 V
Voltage for Gain 200: T = +20 °C: 390.141348 V     T = -25 °C: 354.2535392 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.626629742 V-1     T = -25 °C: 5.005351641 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 9.112465265 V-1     T = -25 °C: 9.325314092 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 14.12071126 V-1     T = -25 °C: 14.67129385 V-1
Break-through voltage: T = +20 °C: 405.9521621 V     T = -25 °C: 369.7007766 V
   
Notes:


Version history

Time Author Change comment
19. May 2022 15:26:31 CEST oafedulidis Assigned to capsule by script
17. May 2021 09:16:18 CEST oafedulidis Updated box and position via API.
05. Feb 2021 11:25:13 CET jreher Updated box and position via API.
06. Oct 2020 18:48:28 CEST jreher Updated APD location via API.
11. Aug 2020 18:04:14 CEST jreher Updated APD characteristics via API.
21. Jul 2020 14:39:57 CEST jreher Updated box and position via API.
10. Feb 2020 18:56:39 CET spankonin Updated APD batch assignment via API.
29. Nov 2019 12:26:27 CET jreher Updated APD location via API.
25. Nov 2019 17:31:10 CET jreher Updated box and position via API.
25. Nov 2019 17:16:22 CET jreher Updated box and position via API.
22. Oct 2019 14:43:18 CEST spankonin
22. Oct 2019 14:21:19 CEST jreher Updated APD batch assignment via API.
20. Aug 2019 14:02:54 CEST jgrochowski Updated APD batch assignment via API.
05. Jul 2019 13:45:28 CEST jreher Updated APD location via API.
03. Jan 2019 00:13:05 CET Tobias Corrected dark current unit.
24. May 2018 14:39:23 CEST Tobias Updated location to "Bochum".
18. May 2018 12:00:45 CEST Tobias Imported from Hamamatsu datasheet.