RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 2410026688 - Archived data from Tue, 11. August 2020 17:22:06 CEST


This is all the information about APD 2410026688. If it is wrong, edit the data.

Subdetector specification:  
Serial: 2410026688
Type: Hamamatsu Avalanche Photo Diode
Detector: assigned to 2410026688/2320025863
Unit: #896 (barcode 1309010103)
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: blue
   
Manufacturer information:  
Wafer position: E08
Break-through voltage: 409 V
Voltage for Gain 100 (T=+25°C): 380.1 V
Dark current: 6.6 nA
   
Screening Logistics:  
Available: No
Storage Box: 14
Position in Box: 44
EP1 batch: none
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: 12. Nov 2018
Sent for analysis after irradiation: 26. Nov 2018
Return for assembly: none
   
Irradiation:  
Date: 22. Nov 2018
Dose used: 37 Gy
Temperature: 20 °C
Position: none
Bias voltage: none     
   
Annealing:  
Date: 23. Nov 2018
Temperature: 80 °C
Duration: 48 h
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 380.2569832 V     T = -25 °C: 344.9152875 V
Voltage for Gain 150: T = +20 °C: 388.1049495 V     T = -25 °C: 352.2966747 V
Voltage for Gain 200: T = +20 °C: 392.5016146 V     T = -25 °C: 356.4366606 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.515843922 V-1     T = -25 °C: 5.001175234 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 9.125838743 V-1     T = -25 °C: 9.042250026 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 15.10874488 V-1     T = -25 °C: 16.04284785 V-1
Break-through voltage: T = +20 °C: 408.3034727 V     T = -25 °C: 371.8326491 V
   
Notes: foam


Version history

Time Author Change comment
05. Feb 2021 10:49:36 CET jreher Updated box and position via API.
03. Feb 2021 18:15:46 CET jreher Updated box and position via API.
11. Aug 2020 17:22:06 CEST jreher Updated APD characteristics via API.
31. Mar 2020 08:51:52 CEST jreher Assigned to capsule by script
13. Feb 2020 14:19:56 CET jreher Updated box and position via API.
27. Nov 2019 19:31:54 CET jreher Updated APD location via API.
25. Nov 2019 17:12:01 CET jreher Updated box and position via API.
04. Jul 2019 11:08:08 CEST jreher Updated APD location via API.
12. Jun 2019 16:46:29 CEST jreher Updated APD batch assignment via API.
12. Jun 2019 16:31:24 CEST jreher Updated APD batch assignment via API.
12. Jun 2019 15:03:43 CEST jreher Updated APD batch assignment via API.
03. Jan 2019 00:13:02 CET Tobias Corrected dark current unit.
26. Nov 2018 12:21:25 CET markus.moritz
16. Nov 2018 12:24:20 CET markus.moritz
12. Jan 2018 13:16:57 CET Tobias Imported from Hamamatsu datasheet.