Show Hamamatsu Avalanche Photo Diode 2408026498 - Archived data from Tue, 11. August 2020 17:21:39 CEST
This is all the information about APD 2408026498. If it is wrong, edit the data.
Subdetector specification: | |
---|---|
Serial: | 2408026498 |
Type: | Hamamatsu Avalanche Photo Diode |
Detector: | unassigned |
Unit: | unassigned |
Preamp: | 0 |
Current location: | Bochum |
Installation information: | |
Label: | none |
Manufacturer information: | |
Wafer position: | D14 |
Break-through voltage: | 415 V |
Voltage for Gain 100 (T=+25°C): | 386.9 V |
Dark current: | 7.5 nA |
Screening Logistics: | |
Available: | Yes |
Storage Box: | 12 |
Position in Box: | 4 |
EP1 batch: | none |
EP1 batch after irradiation: | none |
Shipment: | |
Grid number: | none |
Position in grid: | none |
Arrival for irradiation: | 12. Nov 2018 |
Sent for analysis after irradiation: | 13. Dec 2018 |
Return for assembly: | none |
Irradiation: | |
Date: | 07. Dec 2018 |
Dose used: | 37 Gy |
Temperature: | 20 °C |
Position: | none |
Bias voltage: | none |
Annealing: | |
Date: | 10. Dec 2018 |
Temperature: | 80 °C |
Duration: | 48 h |
Measurement results: | |
Voltage for Gain 100: | T = +20 °C: 387.0368258 V T = -25 °C: 351.0558001 V |
Voltage for Gain 150: | T = +20 °C: 394.8527258 V T = -25 °C: 358.5342713 V |
Voltage for Gain 200: | T = +20 °C: 399.1873465 V T = -25 °C: 362.7075993 V |
Gain/Voltage slope at M = 100: | T = +20 °C: 4.633798732 V-1 T = -25 °C: 4.868852392 V-1 |
Gain/Voltage slope at M = 150: | T = +20 °C: 9.247079 V-1 T = -25 °C: 8.986565861 V-1 |
Gain/Voltage slope at M = 200: | T = +20 °C: 14.49689985 V-1 T = -25 °C: 15.99441878 V-1 |
Break-through voltage: | T = +20 °C: 414.4992251 V T = -25 °C: 378.0486338 V |
Notes: |
Version history
Time | Author | Change comment |
---|---|---|
05. Feb 2021 10:49:16 CET | jreher | Updated box and position via API. |
03. Feb 2021 18:15:24 CET | jreher | Updated box and position via API. |
07. Oct 2020 12:18:17 CEST | jreher | Assigned to capsule by script |
06. Oct 2020 16:51:25 CEST | jreher | Updated box and position via API. |
11. Aug 2020 17:21:39 CEST | jreher | Updated APD characteristics via API. |
27. Nov 2019 19:31:30 CET | jreher | Updated APD location via API. |
25. Nov 2019 17:11:48 CET | jreher | Updated box and position via API. |
04. Jul 2019 11:07:49 CEST | jreher | Updated APD location via API. |
12. Jun 2019 16:46:16 CEST | jreher | Updated APD batch assignment via API. |
12. Jun 2019 16:31:11 CEST | jreher | Updated APD batch assignment via API. |
12. Jun 2019 15:03:00 CEST | jreher | Updated APD batch assignment via API. |
03. Jan 2019 00:12:57 CET | Tobias | Corrected dark current unit. |
13. Dec 2018 14:11:35 CET | markus.moritz | |
30. Nov 2018 13:19:08 CET | markus.moritz | |
12. Jan 2018 13:16:54 CET | Tobias | Imported from Hamamatsu datasheet. |