RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 2405026301


This is all the information about APD 2405026301. If it is wrong, edit the data.

Subdetector specification:  
Serial: 2405026301
Type: Hamamatsu Avalanche Photo Diode
Detector: assigned to 2405026301/2704030291
Unit: #3012 (barcode 1309032853)
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: blue
   
Manufacturer information:  
Wafer position: G12
Break-through voltage: 411 V
Voltage for Gain 100 (T=+25°C): 382.2 V
Dark current: 5.7 nA
   
Screening Logistics:  
Available: No
Storage Box: 9
Position in Box: 47
EP1 batch: none
EP1 batch after irradiation: 10015
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: 12. Nov 2018
Sent for analysis after irradiation: 13. Dec 2018
Return for assembly: none
   
Irradiation:  
Date: 07. Dec 2018
Dose used: 37 Gy
Temperature: 20 °C
Position: none
Bias voltage: none     
   
Annealing:  
Date: 10. Dec 2018
Temperature: 80 °C
Duration: 48 h
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 381.9791093 V     T = -25 °C: 346.7521148 V
Voltage for Gain 150: T = +20 °C: 389.8345013 V     T = -25 °C: 354.16404 V
Voltage for Gain 200: T = +20 °C: 394.2023613 V     T = -25 °C: 358.3301514 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.619555069 V-1     T = -25 °C: 4.629295027 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 9.177496646 V-1     T = -25 °C: 9.357548168 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 14.37120163 V-1     T = -25 °C: 14.7691399 V-1
Break-through voltage: T = +20 °C: 409.0671355 V     T = -25 °C: 373.8276276 V
   
Notes:


Characteristics

Temperature Measurement Notes
No characteristics available!


Upload new characteristic...

Progression of the current during irradiation

Upload time Notes
No data available!


Upload new progression data...

Version history

Time Author Change comment
07. Feb 2022 10:46:00 CET oafedulidis Assigned to capsule by script
17. May 2021 09:17:32 CEST oafedulidis Updated box and position via API.
05. Feb 2021 10:48:58 CET jreher Updated box and position via API.
03. Feb 2021 18:15:06 CET jreher Updated box and position via API.
11. Aug 2020 17:21:17 CEST jreher Updated APD characteristics via API.
27. Nov 2019 19:31:00 CET jreher Updated APD location via API.
25. Nov 2019 17:11:38 CET jreher Updated box and position via API.
04. Jul 2019 11:06:04 CEST jreher Updated APD location via API.
12. Jun 2019 16:46:06 CEST jreher Updated APD batch assignment via API.
12. Jun 2019 16:31:01 CEST jreher Updated APD batch assignment via API.
12. Jun 2019 15:02:27 CEST jreher Updated APD batch assignment via API.
03. Jan 2019 00:12:52 CET Tobias Corrected dark current unit.
13. Dec 2018 14:06:06 CET markus.moritz
23. Nov 2018 11:56:34 CET markus.moritz
12. Jan 2018 13:16:51 CET Tobias Imported from Hamamatsu datasheet.