RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 2404026202 - Archived data from Wed, 03. February 2021 18:14:55 CET


This is all the information about APD 2404026202. If it is wrong, edit the data.

Subdetector specification:  
Serial: 2404026202
Type: Hamamatsu Avalanche Photo Diode
Detector: unassigned
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: none
   
Manufacturer information:  
Wafer position: C10
Break-through voltage: 414 V
Voltage for Gain 100 (T=+25°C): 384.9 V
Dark current: 4.1 nA
   
Screening Logistics:  
Available: Yes
Storage Box: 8
Position in Box: 21
EP1 batch: none
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: 12. Nov 2018
Sent for analysis after irradiation: 26. Nov 2018
Return for assembly: none
   
Irradiation:  
Date: 22. Nov 2018
Dose used: 37 Gy
Temperature: 20 °C
Position: none
Bias voltage: none     
   
Annealing:  
Date: 23. Nov 2018
Temperature: 80 °C
Duration: 48 h
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 384.9531861 V     T = -25 °C: 349.3319202 V
Voltage for Gain 150: T = +20 °C: 392.865449 V     T = -25 °C: 356.7946626 V
Voltage for Gain 200: T = +20 °C: 397.2661475 V     T = -25 °C: 360.9710178 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.608467993 V-1     T = -25 °C: 4.788907414 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 9.129732393 V-1     T = -25 °C: 9.707427946 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 14.2176315 V-1     T = -25 °C: 15.48800519 V-1
Break-through voltage: T = +20 °C: 412.686038 V     T = -25 °C: 376.0857653 V
   
Notes:


Version history

Time Author Change comment
15. Apr 2021 16:58:05 CEST jreher Updated box and position via API.
14. Apr 2021 18:46:23 CEST oafedulidis Assigned to capsule by script
05. Feb 2021 10:48:44 CET jreher Updated box and position via API.
03. Feb 2021 18:14:55 CET jreher Updated box and position via API.
11. Aug 2020 17:21:02 CEST jreher Updated APD characteristics via API.
27. Nov 2019 19:30:47 CET jreher Updated APD location via API.
25. Nov 2019 17:11:31 CET jreher Updated box and position via API.
04. Jul 2019 11:05:53 CEST jreher Updated APD location via API.
12. Jun 2019 16:46:00 CEST jreher Updated APD batch assignment via API.
12. Jun 2019 16:30:55 CEST jreher Updated APD batch assignment via API.
12. Jun 2019 15:01:53 CEST jreher Updated APD batch assignment via API.
03. Jan 2019 00:12:49 CET Tobias Corrected dark current unit.
26. Nov 2018 12:02:48 CET markus.moritz
16. Nov 2018 11:53:26 CET markus.moritz
12. Jan 2018 13:16:49 CET Tobias Imported from Hamamatsu datasheet.