RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 2519029011 - Archived data from Wed, 03. February 2021 18:22:33 CET


This is all the information about APD 2519029011. If it is wrong, edit the data.

Subdetector specification:  
Serial: 2519029011
Type: Hamamatsu Avalanche Photo Diode
Detector: assigned to 2519029011/2514028593
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: blue
   
Manufacturer information:  
Wafer position: B10
Break-through voltage: 410 V
Voltage for Gain 100 (T=+25°C): 381.2 V
Dark current: 3.3 nA
   
Screening Logistics:  
Available: No
Storage Box: 67
Position in Box: 29
EP1 batch: 46
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: 16. Jul 2019
Sent for analysis after irradiation: 19. Jul 2019
Return for assembly: none
   
Irradiation:  
Date: 17. Jul 2019
Dose used: 37 Gy
Temperature: 20 °C
Position: none
Bias voltage: none     
   
Annealing:  
Date: 17. Jul 2019
Temperature: 80 °C
Duration: 48 h
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 381.8232822 V     T = -25 °C: 346.2244573 V
Voltage for Gain 150: T = +20 °C: 389.637993 V     T = -25 °C: 353.640587 V
Voltage for Gain 200: T = +20 °C: 393.978103 V     T = -25 °C: 357.7820785 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.691395072 V-1     T = -25 °C: 4.795761512 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 9.277719689 V-1     T = -25 °C: 9.856752485 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 14.52626188 V-1     T = -25 °C: 15.57147824 V-1
Break-through voltage: T = +20 °C: 408.692675 V     T = -25 °C: 372.3424418 V
   
Notes:


Version history

Time Author Change comment
05. Feb 2021 10:56:59 CET jreher Updated box and position via API.
03. Feb 2021 18:22:33 CET jreher Updated box and position via API.
11. Aug 2020 17:33:33 CEST jreher Updated APD characteristics via API.
03. Apr 2020 09:59:56 CEST jreher Assigned to capsule by script
19. Feb 2020 11:53:34 CET jreher Updated box and position via API.
27. Nov 2019 19:40:39 CET jreher Updated APD location via API.
25. Nov 2019 17:16:12 CET jreher Updated box and position via API.
05. Aug 2019 16:54:42 CEST jreher Updated APD batch assignment via API.
29. Jul 2019 13:21:03 CEST markus.moritz
05. Jul 2019 13:45:13 CEST jreher Updated APD location via API.
01. Jul 2019 12:50:05 CEST jgrochowski Updated APD batch assignment via API.
03. Jan 2019 00:12:37 CET Tobias Corrected dark current unit.