RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 2518028944 - Archived data from Mon, 03. May 2021 10:25:42 CEST


This is all the information about APD 2518028944. If it is wrong, edit the data.

Subdetector specification:  
Serial: 2518028944
Type: Hamamatsu Avalanche Photo Diode
Detector: unassigned
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: none
   
Manufacturer information:  
Wafer position: E08
Break-through voltage: 407 V
Voltage for Gain 100 (T=+25°C): 378.6 V
Dark current: 5.3 nA
   
Screening Logistics:  
Available: Yes
Storage Box: 61
Position in Box: 37
EP1 batch: 44
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: 16. Jul 2019
Sent for analysis after irradiation: 19. Jul 2019
Return for assembly: none
   
Irradiation:  
Date: 16. Jul 2019
Dose used: 37 Gy
Temperature: 20 °C
Position: none
Bias voltage: none     
   
Annealing:  
Date: 17. Jul 2019
Temperature: 80 °C
Duration: 48 h
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 378.3553393 V     T = -25 °C: 343.0481877 V
Voltage for Gain 150: T = +20 °C: 386.2184728 V     T = -25 °C: 350.3951239 V
Voltage for Gain 200: T = +20 °C: 390.5806382 V     T = -25 °C: 354.5377138 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.431431626 V-1     T = -25 °C: 4.894281195 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 8.680382395 V-1     T = -25 °C: 9.007528059 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 14.98793721 V-1     T = -25 °C: 15.97090453 V-1
Break-through voltage: T = +20 °C: 406.2249648 V     T = -25 °C: 369.8211857 V
   
Notes:


Version history

Time Author Change comment
26. Oct 2021 15:12:31 CEST oafedulidis Assigned to capsule by script
03. May 2021 15:20:23 CEST oafedulidis Updated box and position via API.
03. May 2021 10:25:42 CEST oafedulidis Updated box and position via API.
05. Feb 2021 10:56:49 CET jreher Updated box and position via API.
03. Feb 2021 18:22:23 CET jreher Updated box and position via API.
11. Aug 2020 17:33:22 CEST jreher Updated APD characteristics via API.
27. Nov 2019 19:40:26 CET jreher Updated APD location via API.
25. Nov 2019 17:16:06 CET jreher Updated box and position via API.
05. Aug 2019 16:50:14 CEST jreher Updated APD batch assignment via API.
05. Aug 2019 16:50:03 CEST jreher Updated APD batch assignment via API.
29. Jul 2019 11:24:39 CEST markus.moritz
05. Jul 2019 13:45:04 CEST jreher Updated APD location via API.
01. Jul 2019 12:59:50 CEST jgrochowski Updated APD batch assignment via API.
03. Jan 2019 00:12:35 CET Tobias Corrected dark current unit.
12. Jan 2018 13:12:14 CET Tobias Imported from Hamamatsu datasheet.