Show Hamamatsu Avalanche Photo Diode 2517028877 - Archived data from Tue, 11. August 2020 17:33:11 CEST
This is all the information about APD 2517028877. If it is wrong, edit the data.
Subdetector specification: | |
---|---|
Serial: | 2517028877 |
Type: | Hamamatsu Avalanche Photo Diode |
Detector: | unassigned |
Unit: | unassigned |
Preamp: | 0 |
Current location: | Bochum |
Installation information: | |
Label: | none |
Manufacturer information: | |
Wafer position: | E15 |
Break-through voltage: | 406 V |
Voltage for Gain 100 (T=+25°C): | 379.4 V |
Dark current: | 3.9 nA |
Screening Logistics: | |
Available: | Yes |
Storage Box: | 65 |
Position in Box: | 9 |
EP1 batch: | 43 |
EP1 batch after irradiation: | none |
Shipment: | |
Grid number: | none |
Position in grid: | none |
Arrival for irradiation: | 16. Jul 2019 |
Sent for analysis after irradiation: | 19. Jul 2019 |
Return for assembly: | none |
Irradiation: | |
Date: | 16. Jul 2019 |
Dose used: | 37 Gy |
Temperature: | 20 °C |
Position: | none |
Bias voltage: | none |
Annealing: | |
Date: | 17. Jul 2019 |
Temperature: | 80 °C |
Duration: | 48 h |
Measurement results: | |
Voltage for Gain 100: | T = +20 °C: 378.965391 V T = -25 °C: 343.857391 V |
Voltage for Gain 150: | T = +20 °C: 386.8079319 V T = -25 °C: 351.2032264 V |
Voltage for Gain 200: | T = +20 °C: 391.1673664 V T = -25 °C: 355.3176732 V |
Gain/Voltage slope at M = 100: | T = +20 °C: 4.635379691 V-1 T = -25 °C: 4.970031655 V-1 |
Gain/Voltage slope at M = 150: | T = +20 °C: 9.135002739 V-1 T = -25 °C: 9.334957936 V-1 |
Gain/Voltage slope at M = 200: | T = +20 °C: 14.25970145 V-1 T = -25 °C: 14.81323008 V-1 |
Break-through voltage: | T = +20 °C: 404.9061541 V T = -25 °C: 368.9469684 V |
Notes: |
Version history
Time | Author | Change comment |
---|---|---|
05. Feb 2021 10:56:41 CET | jreher | Updated box and position via API. |
03. Feb 2021 18:22:15 CET | jreher | Updated box and position via API. |
11. Aug 2020 17:33:11 CEST | jreher | Updated APD characteristics via API. |
27. Nov 2019 19:40:15 CET | jreher | Updated APD location via API. |
25. Nov 2019 17:16:01 CET | jreher | Updated box and position via API. |
05. Aug 2019 16:48:26 CEST | jreher | Updated APD batch assignment via API. |
25. Jul 2019 15:13:17 CEST | markus.moritz | |
05. Jul 2019 13:44:55 CEST | jreher | Updated APD location via API. |
01. Jul 2019 13:08:18 CEST | jgrochowski | Updated APD batch assignment via API. |
03. Jan 2019 00:12:33 CET | Tobias | Corrected dark current unit. |
12. Jan 2018 13:12:13 CET | Tobias | Imported from Hamamatsu datasheet. |