RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 2517028859 - Archived data from Wed, 03. February 2021 18:22:11 CET


This is all the information about APD 2517028859. If it is wrong, edit the data.

Subdetector specification:  
Serial: 2517028859
Type: Hamamatsu Avalanche Photo Diode
Detector: unassigned
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: none
   
Manufacturer information:  
Wafer position: B06
Break-through voltage: 405 V
Voltage for Gain 100 (T=+25°C): 376.5 V
Dark current: 6.7 nA
   
Screening Logistics:  
Available: Yes
Storage Box: 64
Position in Box: 24
EP1 batch: 42
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: 16. Jul 2019
Sent for analysis after irradiation: 19. Jul 2019
Return for assembly: none
   
Irradiation:  
Date: 16. Jul 2019
Dose used: 37 Gy
Temperature: 20 °C
Position: none
Bias voltage: none     
   
Annealing:  
Date: 17. Jul 2019
Temperature: 80 °C
Duration: 48 h
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 376.9741292 V     T = -25 °C: 341.8974683 V
Voltage for Gain 150: T = +20 °C: 384.7653992 V     T = -25 °C: 349.2170072 V
Voltage for Gain 200: T = +20 °C: 389.1172189 V     T = -25 °C: 353.3311769 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.66169471 V-1     T = -25 °C: 5.012724408 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 9.186673848 V-1     T = -25 °C: 9.363948674 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 14.29103403 V-1     T = -25 °C: 14.80993221 V-1
Break-through voltage: T = +20 °C: 404.372035 V     T = -25 °C: 368.1590121 V
   
Notes:


Version history

Time Author Change comment
29. Mar 2022 12:40:53 CEST oafedulidis Assigned to capsule by script
30. Aug 2021 14:08:26 CEST oafedulidis Updated box and position via API.
03. May 2021 10:25:05 CEST oafedulidis Updated box and position via API.
05. Feb 2021 10:56:37 CET jreher Updated box and position via API.
03. Feb 2021 18:22:11 CET jreher Updated box and position via API.
11. Aug 2020 17:33:08 CEST jreher Updated APD characteristics via API.
27. Nov 2019 19:40:06 CET jreher Updated APD location via API.
25. Nov 2019 17:15:59 CET jreher Updated box and position via API.
05. Aug 2019 16:48:25 CEST jreher Updated APD batch assignment via API.
25. Jul 2019 12:19:32 CEST markus.moritz
05. Jul 2019 13:44:49 CEST jreher Updated APD location via API.
01. Jul 2019 14:31:11 CEST spankonin Updated APD batch assignment via API.
03. Jan 2019 00:12:33 CET Tobias Corrected dark current unit.
12. Jan 2018 13:12:13 CET Tobias Imported from Hamamatsu datasheet.