Show Hamamatsu Avalanche Photo Diode 2515028671 - Archived data from Tue, 11. August 2020 17:32:38 CEST
This is all the information about APD 2515028671. If it is wrong, edit the data.
Subdetector specification: | |
---|---|
Serial: | 2515028671 |
Type: | Hamamatsu Avalanche Photo Diode |
Detector: | unassigned |
Unit: | unassigned |
Preamp: | 0 |
Current location: | Bochum |
Installation information: | |
Label: | none |
Manufacturer information: | |
Wafer position: | B06 |
Break-through voltage: | 411 V |
Voltage for Gain 100 (T=+25°C): | 383 V |
Dark current: | 4.9 nA |
Screening Logistics: | |
Available: | Yes |
Storage Box: | 61 |
Position in Box: | 21 |
EP1 batch: | 37 |
EP1 batch after irradiation: | none |
Shipment: | |
Grid number: | none |
Position in grid: | none |
Arrival for irradiation: | 16. Jul 2019 |
Sent for analysis after irradiation: | 19. Jul 2019 |
Return for assembly: | none |
Irradiation: | |
Date: | 16. Jul 2019 |
Dose used: | 37 Gy |
Temperature: | 20 °C |
Position: | none |
Bias voltage: | none |
Annealing: | |
Date: | 17. Jul 2019 |
Temperature: | 80 °C |
Duration: | 48 h |
Measurement results: | |
Voltage for Gain 100: | T = +20 °C: 383.391237 V T = -25 °C: 347.7330298 V |
Voltage for Gain 150: | T = +20 °C: 391.229412 V T = -25 °C: 355.1223221 V |
Voltage for Gain 200: | T = +20 °C: 395.5887653 V T = -25 °C: 359.2624315 V |
Gain/Voltage slope at M = 100: | T = +20 °C: 4.507532256 V-1 T = -25 °C: 4.685643962 V-1 |
Gain/Voltage slope at M = 150: | T = +20 °C: 8.9088309 V-1 T = -25 °C: 9.521661771 V-1 |
Gain/Voltage slope at M = 200: | T = +20 °C: 14.02814425 V-1 T = -25 °C: 15.32950372 V-1 |
Break-through voltage: | T = +20 °C: 410.7409575 V T = -25 °C: 374.3709782 V |
Notes: |
Version history
Time | Author | Change comment |
---|---|---|
29. Mar 2022 12:38:06 CEST | oafedulidis | Assigned to capsule by script |
27. Aug 2021 10:31:28 CEST | oafedulidis | Updated box and position via API. |
03. May 2021 10:25:40 CEST | oafedulidis | Updated box and position via API. |
05. Feb 2021 10:56:13 CET | jreher | Updated box and position via API. |
03. Feb 2021 18:21:47 CET | jreher | Updated box and position via API. |
11. Aug 2020 17:32:38 CEST | jreher | Updated APD characteristics via API. |
27. Nov 2019 19:39:33 CET | jreher | Updated APD location via API. |
25. Nov 2019 17:15:43 CET | jreher | Updated box and position via API. |
01. Aug 2019 12:12:47 CEST | spankonin | Updated APD batch assignment via API. |
24. Jul 2019 10:28:55 CEST | markus.moritz | |
12. Jul 2019 13:55:40 CEST | jreher | Updated APD location via API. |
01. Jul 2019 14:26:55 CEST | spankonin | Updated APD batch assignment via API. |
03. Jan 2019 00:12:28 CET | Tobias | Corrected dark current unit. |
12. Jan 2018 13:12:10 CET | Tobias | Imported from Hamamatsu datasheet. |