RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 2514028613 - Archived data from Wed, 14. April 2021 18:13:23 CEST


This is all the information about APD 2514028613. If it is wrong, edit the data.

Subdetector specification:  
Serial: 2514028613
Type: Hamamatsu Avalanche Photo Diode
Detector: assigned to 2511028359/2514028613
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: red
   
Manufacturer information:  
Wafer position: C09
Break-through voltage: 409 V
Voltage for Gain 100 (T=+25°C): 379.8 V
Dark current: 6.1 nA
   
Screening Logistics:  
Available: No
Storage Box: 178
Position in Box: 40
EP1 batch: 35
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: 16. Jul 2019
Sent for analysis after irradiation: 19. Jul 2019
Return for assembly: none
   
Irradiation:  
Date: 16. Jul 2019
Dose used: 37 Gy
Temperature: 20 °C
Position: none
Bias voltage: none     
   
Annealing:  
Date: 17. Jul 2019
Temperature: 80 °C
Duration: 48 h
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 380.0699 V     T = -25 °C: 344.5940058 V
Voltage for Gain 150: T = +20 °C: 387.8591413 V     T = -25 °C: 351.9853654 V
Voltage for Gain 200: T = +20 °C: 392.1857325 V     T = -25 °C: 356.1131558 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.660909226 V-1     T = -25 °C: 4.788836142 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 9.284531291 V-1     T = -25 °C: 9.692761913 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 14.57482717 V-1     T = -25 °C: 15.50975199 V-1
Break-through voltage: T = +20 °C: 407.9483256 V     T = -25 °C: 371.3594591 V
   
Notes:


Version history

Time Author Change comment
15. Apr 2021 16:58:00 CEST jreher Updated box and position via API.
14. Apr 2021 18:13:23 CEST oafedulidis Assigned to capsule by script
05. Feb 2021 11:11:17 CET jreher Updated box and position via API.
11. Aug 2020 17:32:28 CEST jreher Updated APD characteristics via API.
27. Nov 2019 20:00:29 CET jreher Updated APD location via API.
25. Nov 2019 17:24:56 CET jreher Updated box and position via API.
30. Jul 2019 15:20:20 CEST spankonin Updated APD batch assignment via API.
24. Jul 2019 09:54:11 CEST markus.moritz
05. Jul 2019 14:50:23 CEST jreher Updated APD location via API.
01. Jul 2019 14:24:18 CEST spankonin Updated APD batch assignment via API.
03. Jan 2019 00:12:26 CET Tobias Corrected dark current unit.
12. Jan 2018 13:12:09 CET Tobias Imported from Hamamatsu datasheet.