RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 2507028004 - Archived data from Tue, 11. August 2020 17:30:57 CEST


This is all the information about APD 2507028004. If it is wrong, edit the data.

Subdetector specification:  
Serial: 2507028004
Type: Hamamatsu Avalanche Photo Diode
Detector: unassigned
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: none
   
Manufacturer information:  
Wafer position: D05
Break-through voltage: 412 V
Voltage for Gain 100 (T=+25°C): 383.3 V
Dark current: 6.8 nA
   
Screening Logistics:  
Available: Yes
Storage Box: 168
Position in Box: 3
EP1 batch: 20
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: 02. Jul 2019
Sent for analysis after irradiation: 09. Jul 2019
Return for assembly: none
   
Irradiation:  
Date: 04. Jul 2019
Dose used: 37 Gy
Temperature: 20 °C
Position: none
Bias voltage: none     
   
Annealing:  
Date: 05. Jul 2019
Temperature: 80 °C
Duration: 48 h
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 383.019767 V     T = -25 °C: 348.3339509 V
Voltage for Gain 150: T = +20 °C: 390.9292441 V     T = -25 °C: 355.5560136 V
Voltage for Gain 200: T = +20 °C: 395.3215046 V     T = -25 °C: 359.5896233 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.654279281 V-1     T = -25 °C: 5.008245282 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 9.308868284 V-1     T = -25 °C: 9.448574059 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 14.73931482 V-1     T = -25 °C: 15.40846551 V-1
Break-through voltage: T = +20 °C: 411.4724616 V     T = -25 °C: 374.6653862 V
   
Notes:


Version history

Time Author Change comment
05. Feb 2021 11:09:55 CET jreher Updated box and position via API.
11. Aug 2020 17:30:57 CEST jreher Updated APD characteristics via API.
27. Nov 2019 19:58:35 CET jreher Updated APD location via API.
25. Nov 2019 17:24:06 CET jreher Updated box and position via API.
30. Jul 2019 15:15:14 CEST spankonin Updated APD batch assignment via API.
09. Jul 2019 15:45:43 CEST markus.moritz
05. Jul 2019 13:58:26 CEST jreher Updated APD location via API.
24. Jun 2019 13:23:06 CEST jgrochowski Updated APD batch assignment via API.
03. Jan 2019 00:12:10 CET Tobias Corrected dark current unit.
12. Jan 2018 13:12:01 CET Tobias Imported from Hamamatsu datasheet.