RUB » Physics » The Faculty » Chairs and Working Groups » EP1 » PANDA FEMC Production Database

Show Hamamatsu Avalanche Photo Diode 2506027983 - Archived data from Tue, 11. August 2020 17:29:58 CEST


This is all the information about APD 2506027983. If it is wrong, edit the data.

Subdetector specification:  
Serial: 2506027983
Type: Hamamatsu Avalanche Photo Diode
Detector: unassigned
Unit: unassigned
Preamp: 0
Current location: Bochum
   
Installation information:  
Label: none
   
Manufacturer information:  
Wafer position: F12
Break-through voltage: 414 V
Voltage for Gain 100 (T=+25°C): 386.4 V
Dark current: 11.5 nA
   
Screening Logistics:  
Available: Yes
Storage Box: 177
Position in Box: 38
EP1 batch: 19
EP1 batch after irradiation: none
   
Shipment:  
Grid number: none
Position in grid: none
Arrival for irradiation: 02. Jul 2019
Sent for analysis after irradiation: 09. Jul 2019
Return for assembly: none
   
Irradiation:  
Date: 04. Jul 2019
Dose used: 37 Gy
Temperature: 20 °C
Position: none
Bias voltage: none     
   
Annealing:  
Date: 05. Jul 2019
Temperature: 80 °C
Duration: 48 h
   
Measurement results:  
Voltage for Gain 100: T = +20 °C: 385.875396 V     T = -25 °C: 350.8334415 V
Voltage for Gain 150: T = +20 °C: 393.7757739 V     T = -25 °C: 358.1849536 V
Voltage for Gain 200: T = +20 °C: 398.1599872 V     T = -25 °C: 362.2828672 V
Gain/Voltage slope at M = 100: T = +20 °C: 4.276721047 V-1     T = -25 °C: 4.678848753 V-1
Gain/Voltage slope at M = 150: T = +20 °C: 8.607816347 V-1     T = -25 °C: 9.847177208 V-1
Gain/Voltage slope at M = 200: T = +20 °C: 15.04593324 V-1     T = -25 °C: 15.87587308 V-1
Break-through voltage: T = +20 °C: 413.5550001 V     T = -25 °C: 377.2369956 V
   
Notes:


Version history

Time Author Change comment
05. Feb 2021 11:11:09 CET jreher Updated box and position via API.
11. Aug 2020 17:29:58 CEST jreher Updated APD characteristics via API.
27. Nov 2019 20:00:17 CET jreher Updated APD location via API.
25. Nov 2019 17:24:51 CET jreher Updated box and position via API.
18. Jul 2019 10:26:29 CEST spankonin Updated APD batch assignment via API.
09. Jul 2019 15:37:56 CEST markus.moritz
05. Jul 2019 14:50:15 CEST jreher Updated APD location via API.
24. Jun 2019 13:18:07 CEST jgrochowski Updated APD batch assignment via API.
03. Jan 2019 00:12:10 CET Tobias Corrected dark current unit.
12. Jan 2018 13:12:01 CET Tobias Imported from Hamamatsu datasheet.